pH ARC Sensors Modbus RTU Programmer’s Manual (EPHUM011)

2.6Configuration of the Measurement

This chapter describes the configuration of PMC1 and PMC6 by means of measurement parameters (PA).

2.6.1 Available Parameters

In register 3072, all available parameters (PA) are given.

Start

Number of

 

Reg1 / Reg2

 

 

 

Modbus

 

Read

Write

register

registers

 

(bitwise defined)

 

 

function code

access

access

3072

2

 

 

Available parameters (see figure 2.6.1.2)

3, 4

 

U/A/S

none

Figure 2.6.1.1: Definition of register 3072.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Bit #

 

Hex value

 

Description

Definition in pH ARC Sensors

 

 

 

 

0 (LSB)

 

0x0001

 

PA1

 

not available

 

 

 

 

 

 

 

 

 

 

 

 

 

not available

 

 

 

 

 

7

 

0x0080

 

PA8

 

not available

 

 

 

 

 

8

 

0x0100

 

PA9

 

Moving average

 

 

 

 

 

9

 

0x0200

 

PA10

 

not available

 

 

 

 

 

10

 

0x0400

 

PA11

 

not available

 

 

 

 

 

11

 

0x0800

 

PA12

 

Moving average R

 

 

 

 

 

 

 

 

 

 

 

 

not available

 

 

 

 

 

15 (MSB)

 

0x8000

 

PA16

 

not available

 

 

 

 

 

Figure 2.6.1.2:

Bitwise definition

of parameters PA1 to PA16, valid for pH ARC Sensors

 

 

 

 

 

 

 

 

 

 

 

 

Command: Available parameters

Modbus address: 3072

Length: 2

Type: 3

 

Read

Parameter:

Measurement

 

 

 

 

 

 

 

 

 

 

 

parameters

 

 

 

 

 

 

 

 

 

Format:

 

Hex

 

 

 

 

 

 

 

 

 

 

 

Value:

 

0x0900

 

 

 

 

 

 

 

 

 

 

 

Figure 2.6.1.3: Example to read the available parameters. The value 0x0900 corresponds to 0x0100 (PA9) + 0x0800 (PA12). Parameter 9 and 12 are available.

Genereal note:

PA1 to PA8 use FLOAT as data format for its values

PA9 to PA16 use UNSIGNED INT as data format for its values.

624300/01

page 26 / 67

Page 26
Image 26
Hamilton Electronics EPHUM011 manual Configuration of the Measurement, Available Parameters