Hamilton Electronics EPHUM011 manual

Models: EPHUM011

1 67
Download 67 pages 49.01 Kb
Page 65
Image 65

pH ARC Sensors Modbus RTU Programmer’s Manual (EPHUM011)

Command: Serial number

Modbus address: 1312

Length: 8

Type: 3

Read

Parameter:

Text

 

 

 

 

Format:

Character

 

 

 

 

Value:

0001001

 

 

 

 

Figure 2.9.2.2: Example to read register 1312.

2.9.3 Free User Memory Space

These registers can be used to store any customer specific information in the sensor. There are different registers which can be read by everybody, but only specific operators can write them.

Start

Number

Reg1 … Reg8

Example of content

Modbus

Read

Write

register

of

(16 ASCII characters)

 

function

access

access

 

registers

 

 

code

 

 

1536

8

Free user space U/A/S

*FREE_USERSPACE*

3, 4, 16

U/A/S

U/A/S

1544

8

Free user space U/A/S

*FREE_USERSPACE*

3, 4, 16

U/A/S

U/A/S

1552

8

Free user space U/A/S

*FREE_USERSPACE*

3, 4, 16

U/A/S

U/A/S

1560

8

Free user space U/A/S

*FREE_USERSPACE*

3, 4, 16

U/A/S

U/A/S

1568

8

Free user space A/S

*FREE_USERSPACE*

3, 4, 16

U/A/S

A/S

1576

8

Free user space A/S

*FREE_USERSPACE*

3, 4, 16

U/A/S

A/S

1584

8

Free user space A/S

*FREE_USERSPACE*

3, 4, 16

U/A/S

A/S

1592

8

Free user space A/S

*FREE_USERSPACE*

3, 4, 16

U/A/S

A/S

1600

8

Measuring point

242111-0001001

3, 4, 16

U/A/S

S

1608

8

Free user space S

*FREE_USERSPACE*

3, 4, 16

U/A/S

S

1616

8

Free user space S

*FREE_USERSPACE*

3, 4, 16

U/A/S

S

1624

8

Free user space S

*FREE_USERSPACE*

3, 4, 16

U/A/S

S

1632

8

Free user space others

*FREE_USERSPACE*

3, 4

U/A/S

none

1640

8

Free user space others

*FREE_USERSPACE*

3, 4

U/A/S

none

1648

8

Free user space others

*FREE_USERSPACE*

3, 4

U/A/S

none

1656

8

Free user space others

*FREE_USERSPACE*

3, 4

U/A/S

none

1664

8

Free user space others

*FREE_USERSPACE*

3, 4

U/A/S

none

1672

8

Free user space others

*FREE_USERSPACE*

3, 4

U/A/S

none

1680

8

Free user space others

*FREE_USERSPACE*

3, 4

U/A/S

none

1688

8

Free user space others

*FREE_USERSPACE*

3, 4

U/A/S

none

1696

8

Free user space others

*FREE_USERSPACE*

3, 4

U/A/S

none

1704

8

Free user space others

*FREE_USERSPACE*

3, 4

U/A/S

none

1712

8

Free user space others

*FREE_USERSPACE*

3, 4

U/A/S

none

1720

8

Free user space others

*FREE_USERSPACE*

3, 4

U/A/S

none

1728

8

Free user space others

*FREE_USERSPACE*

3, 4

U/A/S

none

1736

8

Free user space others

*FREE_USERSPACE*

3, 4

U/A/S

none

1744

8

Free user space others

*FREE_USERSPACE*

3, 4

U/A/S

none

1752

8

Free user space others

*FREE_USERSPACE*

3, 4

U/A/S

none

Figure 2.9.3.1: Definition of registers containing user information.

An important register is 1600, as it is the description of the measuring point. The information of this register is displayed on the ARC View Handheld in order to identify individual sensors.

Attention:

The Free User Memory Space is located in a memory which allows in total max 10’000 write operations.

624300/01

page 65 / 67

Page 65
Image 65
Hamilton Electronics EPHUM011 manual