Hamilton Electronics EPHUM011 manual Description of PA9 Moving Average

Models: EPHUM011

1 67
Download 67 pages 49.01 Kb
Page 28
Image 28

pH ARC Sensors Modbus RTU Programmer’s Manual (EPHUM011)

2.6.2.1Description of PA9 (Moving Average)

In register 3360, a plain text ASCII description of PA9 is given.

Start

Number of

Reg1 … Reg8

Modbus

Read

Write

register

registers

16 ASCII characters

function code

access

access

3360

8

Description of PA9

3, 4

U/A/S

none

Figure 2.6.2.1.1: Definition of register 3360.

Command: Moving average text

Modbus address: 3360

Length: 8

Type: 3

Read

Parameter:

Text

 

 

 

 

 

 

Format:

Character

 

 

 

 

 

 

Value:

Moving average

 

 

 

 

 

 

Figure 2.6.2.1.2: Example to read the description for “Moving average“.

2.6.2.2Selecting the Physical Unit and Writing the Value for PA9

In register 3368, the available physical units for PA9 are defined.

Start

Number of

Reg1 / Reg2

Modbus

 

Read

Write

register

registers

(bitwise defined)

function code

 

access

access

3368

2

 

Available physical units for PA9

3, 4

 

U/A/S

none

Figure 2.6.2.2.1: Definition of register 3368.

 

 

 

 

 

 

 

 

 

 

 

 

Command: Moving average av. units

Modbus address: 3368

Length: 2

Type: 3

Read

Parameter:

Units

 

 

 

 

 

 

 

 

Format:

 

Hex

 

 

 

 

 

 

 

 

Value:

 

0x01

 

 

 

 

 

 

 

 

Figure 2.6.2.2.2: Example to read the available physical units for PA9. The only one available here is “none”(0x01). For the definition of the physical units see chapter 2.5.1.

Start

Number of

Reg1 / Reg2

Reg3 / Reg4

Modbus

Read

Write

register

registers

(bitwise defined)

 

function code

access

access

3370

4

Select physical unit

Value for PA9

16

none

S

 

 

for PA9

(1-16, default: 2)

 

 

 

Figure 2.6.2.2.3: Definition of register 3370. Only one bit for the physical unit can be set. PA9 can be set to the value 1-16. A value of 1 does not influence the response time of the sensor, a value of 16 increases the response time of the sensor to 48 s.

By writing to register 3370 the active physical unit for PA9 can be selected by choosing one of the physical units that are defined in register 3368. The value of the parameter can be set as well.

Command: Moving average

Modbus address: 3370

Length: 4

Type: 16

Write

Parameter:

Unit

Value

 

 

 

 

Format:

Hex

Decimal

 

 

 

 

Value:

0x01

12

 

 

 

 

Figure 2.6.2.2.4: Example to set the physical unit of PA9 to “none”(0x01) and the value of the moving average to 12.

624300/01

page 28 / 67

Page 28
Image 28
Hamilton Electronics EPHUM011 Description of PA9 Moving Average, Selecting the Physical Unit and Writing the Value for PA9