pH ARC Sensors Modbus RTU Programmer’s Manual (EPHUM011)

2.4.5 Configuration of the 4-20 mA Interface

Note:

The configuration of AO1 / AO2 is only effective if register 4360 / 4488 (active analog interface mode) is set to the value 0x01, 0x02 or 0x04.

2.4.5.1Reading the Available Primary Measurement Channels to be Mapped to the Analog Output

Start

Number of

Reg1 / Reg2

Modbus

Read

Write

register

registers

 

function code

access

access

4362

2

Available Primary Measurement Channels

3, 4

U/A/S

none

 

 

for AO1

 

 

 

4490

2

Available Primary Measurement Channels

3, 4

U/A/S

none

 

 

for AO2

 

 

 

Figure 2.4.5.1.1: Definition of register 4362 / 4490.

For the definition of the Primary Measurement Channels (PMC), see chapter 2.5.

Primary Measurement Channel (PMC)

PMC1 (pH)

not available

0x20 PMC6 (temperature)

Figure 2.4.5.1.2: Code for selection of the primary measurement channel.

Command: Available PMC AO1

Modbus address: 4362

Length: 2

Type: 3

Read

Parameter:

Available PMC 20 mA

 

 

 

 

 

 

Format:

hex

 

 

 

 

 

 

Value:

0x21

 

 

 

 

 

 

Figure 2.4.5.1.3: Example to read the available Primary Measurement Channels (PMC) for AO1. The hexadecimal value of “0x21”defines that PMC1 (pH) or PMC6 (temperature) can be mapped to AO1. Register 4490 contains the same value “0x21”. Accordingly, PMC1 or PMC6 can be mapped to AO2 as well.

2.4.5.2Selecting the Primary Measurement Channel to be Mapped to the Analog Interface

Start

Number of

Reg1 / Reg2

Modbus

Read

Write

register

registers

 

function code

access

access

4364

2

Selected PMC for AO1

3, 4, 16

U/A/S

S

4492

2

Selected PMC for AO2

3, 4, 16

U/A/S

S

Figure 2.4.5.2.1: Definition of register 4364 / 4492. Only one bit can be set.

Command: Active PMC AO1

Modbus address: 4364

Length: 2

Type: 3

Read

Parameter:

Current PMC 20mA

 

 

 

 

 

 

Format:

hex

 

 

 

 

 

 

Value:

0x01

 

 

 

 

 

 

Figure 2.4.5.2.2: Example to read the current primary measurement channel mapped to AO1, defined in register 4364. The value “0x01”is returned, saying that PMC1 is mapped to AO1 (factory setting).

The factory setting for register 4492 is “0x20”, mapping PMC6 to AO2.