pH ARC Sensors Modbus RTU Programmer’s Manual (EPHUM011)

Command: Calibration status CP1

Modbus address: 5158

Length: 6

Type: 3

Read

Parameter:

Status

Unit

Value

 

 

 

Format:

Hex

Hex

Float

 

 

 

Value:

0x00000080

0x00001000

4.01

 

 

 

Figure 2.7.4.1.3: Example to read the calibration status of CP1 after calibration CP1 at 0 = auto. The status message is: “CP1 pH reading during calibration is not stable”(0x00000080). The physical unit of the last successful calibration is pH (0x00001000) and the last successful calibration has been performed at pH 4.01.

Command: Calibration status CP2

Modbus address: 5190

Length: 6

Type: 3

Read

Parameter:

Status

Unit

Value

 

 

 

Format:

Hex

Hex

Float

 

 

 

Value:

0x00000080

0x00001000

7

 

 

 

Figure 2.7.4.1.4: Example to read the calibration status of CP2 after calibrating CP2 at 0 = auto. All bits of CP2 are zero (0x00000080), indicating that the calibration was successful. The physical unit of the last calibration is pH (0x00001000) and the last successful calibration has been performed at pH 7.

Command: Calibration status CP2

Modbus address: 5190

Length: 6

Type: 3

Read

Parameter:

Status

Unit

Value

 

 

 

Format:

Hex

Hex

Float

 

 

 

Value:

0x00000180

0x00001000

7

 

 

 

Figure 2.7.4.1.5: Example to read the calibration status of CP2 after attempt to calibrate CP2 at pH 4.01, which is the same as CP1. The value is 0x00000180 = 0x00000080 + 0x00000100. Shown is still 0x00000080 of CP1 and new 0x00000100 of CP2, which says: “CP2 space to CP1 < pH 1.0”.

2.7.4.2Reading the Calibration Status of CP6 (Product Calibration)

The calibration status and the current state of the product calibration process (CP6) is read in the calibration status register for CP6 (register 5318).

Start

Number of

Reg1 / Reg2

Reg3 / Reg4

Reg5 / Reg6

Modbus

Read

Write

register

registers

 

 

 

function

access

access

 

 

 

 

 

code

 

 

5318

6

Status CP6

Physical unit of

pH value of the

3, 4

U/A/S

none

 

 

(see figure

the last

last successful

 

 

 

 

 

2.7.4.1.1)

successful

calibration CP6

 

 

 

 

 

 

calibration CP6

 

 

 

 

 

 

 

(always pH)

 

 

 

 

Figure 2.7.4.2.1: Definition of register 5318 for CP6. For examples, see following chapters.

624300/01

page 45 / 67

Page 45
Image 45