pH ARC Sensors Modbus RTU Programmer’s Manual (EPHUM011)

2.4.5.8Defining the Error and Warning Output of the 4-20 mA Interface

Errors and warnings can be mapped to the AO1 / AO2.

Start

Number of

Reg1 /

Reg3 /

Reg5 /

Reg7 /

Modbus

Read

Write

register

registers

Reg2

Reg4

Reg6

Reg8

function

access

access

 

 

 

 

 

 

code

 

 

4386

8

Code of

Current

Current

Current in

3, 4, 16

U/A/S

S

 

 

warnings

in case of

in case of

case of

 

 

 

 

 

and errors

“warning”

“error”

“T exceed”

 

 

 

 

 

(see Figure

[mA]

[mA]

[mA]

 

 

 

 

 

2.4.4.8.2)

for AO1

for AO1

for AO1

 

 

 

 

 

for AO1

 

 

 

 

 

 

4514

8

Code of

Current

Current

Current in

3, 4, 16

U/A/S

S

 

 

warnings

in case of

in case of

case of

 

 

 

 

 

and errors

“warning”

“error”

“T exceed”

 

 

 

 

 

(see Figure

[mA]

[mA]

[mA]

 

 

 

 

 

2.4.4.8.2)

for AO2

for AO2

for AO2

 

 

 

 

 

for AO2

 

 

 

 

 

 

Figure 2.4.5.8.1: Definition of register 4386 / 4514.

Bit #

Code (hex)

Behaviour of the 4-20 mA interface in case of errors and warnings

0 (LSB)

0x000001

Error continuous output

 

 

not available

16

0x010000

Warning continuous output

 

 

not available

Figure 2.4.5.8.2: Code for the 4-20 mA interface in case of errors and warnings.

If the corresponding bits for the errors and warnings are not set (=0x00), the respective options are inactive.

The default settings are:

Code 0x01

current in case of warnings:

3.5 mA

current in case of errors:

3.5 mA

current in case of temperature exceed: 3.5 mA

Command: ErrorWarnings AO1

Modbus address: 4386

Length: 8

Type: 3

Read

Parameter:

Warning code

Current in case of

Current

in case of

Current in case of

 

 

warning

error

 

temperature exceed

 

 

[mA]

[mA]

 

[mA]

 

Format:

Hex

Float

Float

 

Float

 

Value:

0x010001

3.5

3.5

 

3.5

 

Figure 2.4.5.8.4: Example: Read the settings for AO1 in case of warnings and errors. Warning code 0x010001 corresponds to the continuous output current in case of warning (0x010000) and continuous output current in case of error (0x01) of 3.5 mA. The output current in case of temperature exceed is 3.5 mA.

624300/01

page 16 / 67

Page 16
Image 16
Hamilton Electronics EPHUM011 manual 0x010001