pH ARC Sensors Modbus RTU Programmer’s Manual (EPHUM011)

2.8Sensor Status

2.8.1 Temperature Ranges

In registers 4608, 4612 and 4616 three different temperature ranges are defined:

Operation –in this range the sensor will work properly (current output, Modbus communication), except the measurement, which is stopped until the temperature is back in the measurement range. In this case the last value of measurement will be frozen and sent to analog interfaces.

Measurement –in this range the sensor is able to measure.

Calibration –in this range the sensor can be calibrated.

Start

Number of

Reg1 / Reg2

Reg3 / Reg4

 

Modbus

Read

Write

register

registers

 

 

 

function

access

access

 

 

 

 

 

code

 

 

4608

4

Operating

Operating

 

3, 4

U/A/S

none

 

 

temperature min [°C]

temperature max

[°C]

 

 

 

4612

4

Measurement

Measurement

 

3, 4

U/A/S

none

 

 

temperature min [°C]

temperature max [°C]

 

 

 

4616

4

Calibration

Calibration

 

3, 4

U/A/S

none

 

 

temperature min [°C]

temperature max [°C]

 

 

 

Figure 2.8.1.1: Definition of register 4608, 4612 and 4616.

Command: Operating T range

 

Modbus address: 4608

Length: 4

Type: 3

Read

Parameter:

Operating T min

Operating T max

 

 

 

 

 

[°C]

[°C]

 

 

 

 

Format:

Float

Float

 

 

 

 

Value:

-20

130

 

 

 

 

Figure 2.8.1.2:

Example to read the

operating temperature

values min and max.

 

 

 

 

 

 

 

 

Command: Measurement T range

 

Modbus address: 4612

Length: 4

Type: 3

Read

Parameter:

Measurement T

Measurement T

 

 

 

 

 

min [°C]

max [°C]

 

 

 

 

Format:

Float

Float

 

 

 

 

Value:

-20

130

 

 

 

 

Figure 2.8.1.3:

Example to read the

measurement temperature values min and max.

 

 

 

 

 

 

 

Command: Calibration T range

 

Modbus address: 4616

Length: 4

Type: 3

Read

Parameter:

Calibration T min

Calibration T max

 

 

 

 

 

[°C]

[°C]

 

 

 

 

Format:

Float

Float

 

 

 

 

Value:

5

50

 

 

 

 

 

Figure 2.8.1.4: Example to read the calibration temperature values min and max.