pH ARC Sensors Modbus RTU Programmer’s Manual (EPHUM011)

2.8.2 Operating Hours and Counters

In register 4676 are given:

total operating hours

operating hours above max measurement temperature (see chapter 2.8.1)

the operating hours above max operating temperature (see chapter 2.8.1)

In register 4682 are stored:

number of power ups

number of watchdog resets

number of writing cycles to the sensor’s flash memory

In register 4688 are given:

number of sterilizations in place (SIP) (see chapter 2.8.5)

number of cleanings in place (CIP) (see chapter 2.8.5)

Start

Number of

Reg1 / Reg2

Reg3 / Reg4

Reg3 / Reg4

Modbus

Read

Write

register

registers

 

 

 

function code

access

access

4676

6

Operating

Operating

Operating

3, 4

U/A/S

none

 

 

hours

hours above

hours above

 

 

 

 

 

[h]

max

max

 

 

 

 

 

 

measurement

operating

 

 

 

 

 

 

temperature

temperature

 

 

 

 

 

 

[h]

[h]

 

 

 

4682

6

Number of

Number of

Number of

3, 4

U/A/S

none

 

 

Power ups

Watchdog

Writing

 

 

 

 

 

 

resets

cycles to

 

 

 

 

 

 

 

flash memory

 

 

 

4688

4

Number of

Number of

-

3, 4

U/A/S

none

 

 

SIP cycles

CIP cycles

 

 

 

 

Figure 2.8.2.1: Definition of register 4676, 4682 and 4688.

Command: Operating hours

 

Modbus address: 4676

Length: 6

Type: 3

Read

Parameter:

Operating hours [h]

 

Operating hours

Operating hours

 

 

 

 

above max

above max

 

 

 

 

measurement

operating

 

 

 

 

temperature

temperature

 

 

 

 

[h]

[h]

 

 

 

Format:

Float

Float

Float

 

 

 

Value:

168.3667

0

 

0

 

 

 

Figure 2.8.2.2: Example to read the total operating hours, the operating hours above the max measurement temperature and the operating hours above the max operating temperature.

Command: Power & watchdog

 

Modbus address: 4682

 

Length: 6

Type: 3

Read

Parameter:

Number of Power

Number of

Number

of Writing

 

 

 

ups

Watchdog resets

cycles to flash

 

 

 

 

 

 

memory

 

 

 

Format:

Decimal

Decimal

Decimal

 

 

 

Value:

34

1

 

16

 

 

 

 

Figure 2.8.2.3: Example to read the number of power ups, the number of watchdog resets and the number of writing cycles to flash memory.

Command: SIP & CIP

 

Modbus address: 4688

Length: 4

Type: 3

Read

Parameter:

SIP cycles

CIP cycles

 

 

 

 

Format:

Decimal

Decimal

 

 

 

 

Value:

0

0

 

 

 

 

 

Figure 2.8.2.4: Example to read the number of SIP cycles and the number of CIP cycles. For the definition of SIP and CIP cycles see chapter 2.8.5.

624300/01

page 58 / 67

Page 58
Image 58
Hamilton Electronics EPHUM011 manual Operating Hours and Counters, 168.3667