pH ARC Sensors Modbus RTU Programmer’s Manual (EPHUM011)

Command: Read calibration stability

Modbus address: 5128

Length: 4

Type: 3

Read

Parameter:

Max drift pH

 

Max drift Temp

 

 

 

 

[pH/min]

 

[K/min]

 

 

 

Format:

Float

 

Float

 

 

 

Value:

0.1

 

0.5

 

 

 

Figure 2.7.2.1.5:

Example to read the

calibration stability.

 

 

 

 

 

 

 

 

Command: Set calibration stability

Modbus address: 5128

Length: 4

Type: 16

Write

Parameter:

Max drift pH

 

Max drift Temp

 

 

 

 

[pH/min]

 

[K/min]

 

 

 

Format:

Float

 

Float

 

 

 

Value:

0.2

 

0.5

 

 

 

Figure 2.7.2.1.6: Example to set the calibration stability.

Attention:

The stability criteria defined in register 5128 is valid for CP1 and CP2 only, but NOT for CP6.

2.7.2.2Calibration Point 6 (Product Calibration)

The limits for calibration point 6 are given in register 5312.

Start

Number of

Reg1 / Reg2

Reg3 / Reg4

Reg5 / Reg6

Modbus

Read

Write

register

registers

 

 

 

function code

access

access

5312

6

Physical unit

Min value for

Max value for

3, 4

U/A/S

none

 

 

currently

CP6 (in the

CP6 (in the

 

 

 

 

 

active for

physical unit

physical unit

 

 

 

 

 

CP6

as defined in

as defined in

 

 

 

 

 

 

Reg1 and 2)

Reg1 and 2)

 

 

 

Figure 2.7.2.2.1: Definition of register 5312 for CP6.

Command: Calibration limits CP6

 

Modbus address: 5312

Length: 6

Type: 3

Read

Parameter:

Unit

Min value

Max value

 

 

Format:

Hex

Float

Float

 

 

 

Value:

0x01000

0

 

14

 

 

 

Figure 2.7.2.2.2: Example to read the limits of CP6. The active physical unit is pH, the min value is pH 0 and the max value is pH 14.

Note: the definition of min and max is different than the one for CP1 / CP2, because CP6 can be set to any pH value.

624300/01

page 34 / 67