pH ARC Sensors Modbus RTU Programmer’s Manual (EPHUM011)

2.7Calibration

2.7.1 Available Calibration Points

In register 5120, the available number of Calibration Points (CP) for Primary Measurement Channel 1 (PMC1) is defined. 8 individual CP are theoretically possible.

Start

Number of

Reg1 / Reg2

Modbus

Read

Write

register

registers

(bitwise defined)

function code

access

access

5120

2

Available number of CP for PMC1

3, 4

U/A/S

none

 

 

(see figure 2.7.1.2)

 

 

 

Figure 2.7.1.1: Definition of register 5120.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Bit #

Hex value

 

Description

Definition in ARC Sensors

 

 

0

(LSB)

0x01

 

CP1

Calibration Point 1

 

 

 

1

 

0x02

 

CP2

Calibration Point 2

 

 

 

2

 

0x04

 

CP3

not available

 

 

 

 

not available

 

 

 

5

 

0x20

 

CP6

Product Calibration

 

 

 

6

 

0x40

 

CP7

not available

 

 

 

7

(MSB)

0x80

 

CP8

not available

 

 

 

Figure 2.7.1.2:

Bitwise definition

of CP1 to CP8.

 

 

 

 

 

 

 

 

 

 

 

Command: Available cali points

 

Modbus address: 5120

Length: 2

Type: 3

Read

Parameter:

Points

 

 

 

 

 

 

 

Format:

Hex

 

 

 

 

 

 

 

Value:

0x23

 

 

 

 

 

 

 

Figure 2.7.1.3: Example to read the available CPs. 0x23 = 0x01 (CP1) + 0x02 (CP2) + 0x20 (CP6).

 

 

 

Standard Calibration at 2 calibration points

 

 

 

 

200

 

 

 

 

 

 

 

 

 

 

 

 

 

 

ref [mV]

150

 

 

 

 

CP1

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

100

 

 

 

 

 

 

 

 

 

 

 

 

 

 

vs.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

glass

50

 

 

 

 

 

 

 

CP2

 

 

 

 

 

 

0

 

 

 

 

 

 

 

 

 

 

 

 

 

 

potential

 

 

 

 

 

 

 

 

 

 

 

 

 

 

-50

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

electrical

-100

 

 

 

 

 

 

 

 

 

 

 

 

 

 

-150

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

-200

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

0

1

2

3

4

5

6

7

8

9

10

11

12

13

14

 

 

 

 

 

 

 

 

pH value

 

 

 

 

 

 

Figure 2.7.1.4: pH ARC Sensors allow 3 calibration points:

CP1 and CP2 are used for standard calibration (shown in this figure).

The product calibration CP6 is used to adjust the standard calibration function to specific process conditions (the effect of CP6 is shown in Figure 2.7.3.2.1).

624300/01

page 32 / 67

Page 32
Image 32