pH ARC Sensors Modbus RTU Programmer’s Manual (EPHUM011)
2.7Calibration
2.7.1 Available Calibration Points
In register 5120, the available number of Calibration Points (CP) for Primary Measurement Channel 1 (PMC1) is defined. 8 individual CP are theoretically possible.
Start | Number of | Reg1 / Reg2 | Modbus | Read | Write |
register | registers | (bitwise defined) | function code | access | access |
5120 | 2 | Available number of CP for PMC1 | 3, 4 | U/A/S | none |
|
| (see figure 2.7.1.2) |
|
|
|
Figure 2.7.1.1: Definition of register 5120. |
|
|
|
|
| |||||
|
|
|
|
|
|
|
|
|
| |
Bit # | Hex value |
| Description | Definition in ARC Sensors |
|
| ||||
0 | (LSB) | 0x01 |
| CP1 | Calibration Point 1 |
|
|
| ||
1 |
| 0x02 |
| CP2 | Calibration Point 2 |
|
|
| ||
2 |
| 0x04 |
| CP3 | not available |
|
|
| ||
… | … |
| … | not available |
|
|
| |||
5 |
| 0x20 |
| CP6 | Product Calibration |
|
|
| ||
6 |
| 0x40 |
| CP7 | not available |
|
|
| ||
7 | (MSB) | 0x80 |
| CP8 | not available |
|
|
| ||
Figure 2.7.1.2: | Bitwise definition | of CP1 to CP8. |
|
|
| |||||
|
|
|
|
|
|
|
| |||
Command: Available cali points |
| Modbus address: 5120 | Length: 2 | Type: 3 | Read | |||||
Parameter: | Points |
|
|
|
|
|
|
| ||
Format: | Hex |
|
|
|
|
|
|
| ||
Value: | 0x23 |
|
|
|
|
|
|
|
Figure 2.7.1.3: Example to read the available CPs. 0x23 = 0x01 (CP1) + 0x02 (CP2) + 0x20 (CP6). | |||||||||||||||
|
|
| Standard Calibration at 2 calibration points |
|
|
| |||||||||
| 200 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
ref [mV] | 150 |
|
|
|
| CP1 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| ||
100 |
|
|
|
|
|
|
|
|
|
|
|
|
|
| |
vs. |
|
|
|
|
|
|
|
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |
glass | 50 |
|
|
|
|
|
|
| CP2 |
|
|
|
|
|
|
0 |
|
|
|
|
|
|
|
|
|
|
|
|
|
| |
potential |
|
|
|
|
|
|
|
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
|
|
|
|
| ||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |
electrical |
|
|
|
|
|
|
|
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
|
|
|
|
| ||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |
| 0 | 1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 | 9 | 10 | 11 | 12 | 13 | 14 |
|
|
|
|
|
|
|
| pH value |
|
|
|
|
|
|
Figure 2.7.1.4: pH ARC Sensors allow 3 calibration points:
CP1 and CP2 are used for standard calibration (shown in this figure).
The product calibration CP6 is used to adjust the standard calibration function to specific process conditions (the effect of CP6 is shown in Figure 2.7.3.2.1).
624300/01 | page 32 / 67 |