pH ARC Sensors Modbus RTU Programmer’s Manual (EPHUM011)

2.7.2 Definitions of Calibration Points

2.7.2.1Calibration Points 1 and 2 (Standard Calibration)

The limits for the calibration point 1 are defined in register 5152, for calibration point 2 in register 5184.

Start

Number of

Reg1 / Reg2

Reg3 / Reg4

Reg5 / Reg6

Modbus

Read

Write

register

registers

 

 

 

function code

access

access

5152

6

Physical unit

Min value for

Max value for

3, 4

U/A/S

none

 

 

currently

CP1 (in the

CP1 (in the

 

 

 

 

 

active for

physical unit

physical unit

 

 

 

 

 

CP1

as defined in

as defined in

 

 

 

 

 

 

Reg1 and 2)

Reg1 and 2)

 

 

 

5184

6

Physical unit

Min value for

Max value for

3, 4

U/A/S

none

 

 

currently

CP2 (in the

CP2 (in the

 

 

 

 

 

active for

physical unit

physical unit

 

 

 

 

 

CP2

as defined in

as defined in

 

 

 

 

 

 

Reg1 and 2)

Reg1 and 2)

 

 

 

Figure 2.7.2.1.1: Definition of register 5152 for CP1 and 5184 for CP2.

Attention:

The only physical unit available for calibration is pH ! The physical unit defined in 5152, 5184 and 5312 for CP1, CP2 and CP6 is NOT linked to the physical unit defined for PMC1 in register 2090.

When performing a calibration while having the physical unit set to “mV”, the calibration status will report “CP1: incorrect measurement unit”or “CP2: incorrect measurement unit”(see figure 2.7.4.1.1).

Command: Calibration limits CP1

 

Modbus address: 5152

Length: 6

Type: 3

Read

Parameter:

Unit

Min value

Max value

 

 

Format:

Hex

Float

Float

 

 

 

Value:

0x01000

0

 

0

 

 

 

Figure 2.7.2.1.2: Example to read the limits of CP1. Currently active physical unit is pH (0x01000). The min and max values are both 0, indicating, that calibration at CP1 can be performed only using defined calibration standards having discrete pH values.

Command: Calibration limits CP2

 

Modbus address: 5184

Length: 6

Type: 3

Read

Parameter:

Unit

Min value

Max value

 

 

Format:

Hex

Float

Float

 

 

 

Value:

0x01000

0

 

0

 

 

 

Figure 2.7.2.1.3: Example to read the limits of CP2. The active physical unit is is pH (0x01000). The min and max values are both 0, indicating, that calibration at CP2 can be performed only using defined calibration standards having discrete pH values.

When initiating the calibration at CP1 and CP2, the measured pH and temperature have to be stable for at least 3 minutes. The stability criteria are defined in register 5128:

Start

Number of

Reg1 / Reg2

Reg3 / Reg4

Modbus

Read

Write

register

registers

(Float)

(Float)

function code

access

access

5128

4

Max. Drift PMC1

Max. Drift PMC6

3, 4, 16

U/A/S

S

 

 

pH

Temperature

 

 

 

 

 

[pH/min]

[K/min]

 

 

 

Figure 2.7.2.1.4: Definition of register 5128.

624300/01

page 33 / 67

Page 33
Image 33
Hamilton Electronics EPHUM011 Definitions of Calibration Points, Calibration Points 1 and 2 Standard Calibration, CP1, CP2