pH ARC Sensors Modbus RTU Programmer’s Manual (EPHUM011)

Command: Operator level

 

Modbus address: 4288

Length: 4

Type: 3

Read

Parameter:

Operator level

 

Password

 

 

 

 

Format:

Hex

decimal

 

 

 

 

Value:

0x30

0

 

 

 

 

 

Figure 2.2.1.4: Example to read the active operator level: the active level is 0x30 (Specialist). The sensor does not report the password. The value 0 is returned instead.

Command: Operator level

 

Modbus address: 4288

Length: 4

Type: 16

Write

Parameter:

Operator level

 

Password

 

 

 

 

Format:

Hex

decimal

 

 

 

 

Value:

0x03

0

 

 

 

 

 

Figure 2.2.1.5:

Example to set the operator level to 0x03 (User). The password 0 has to be sent.

 

 

 

 

 

 

 

Command: Operator level

 

Modbus address: 4288

Length: 4

Type: 16

Write

Parameter:

Operator level

 

Password

 

 

 

 

Format:

Hex

decimal

 

 

 

 

Value:

0x0C

18111978

 

 

 

 

Figure 2.2.1.6: Example to set the active operator level to 0xC (Administrator). The correct password has to be sent.

Command: Operator level

Modbus address: 4288

Length: 4

Type: 16

Write

Parameter:

Operator level

Password

 

 

 

 

Format:

Hex

decimal

 

 

 

 

Value:

0x0B

18111978

 

 

 

 

Figure 2.2.1.7: Example for a Modbus error. If the level or the password is not correct, (Operator level = 0x0B), the sensor answers with a Modbus error message “Slave device exception response” (see chapter 1.6).

2.2.2 Changing Passwords for Operator Level

The passwords for accessing the operator levels A and S can be modified by S (Specialist) only. U (User) and A (Administrator) have no right to change any password. The new password will remain stored after power down.

Start

 

Number of

Reg1 / Reg2

Reg3 / Reg4

Modbus

 

Read

Write

 

register

 

registers

 

 

 

 

function code

 

access

access

 

4292

 

4

 

Level

 

New password

16

 

 

None

S

 

Figure 2.2.2.1: Definition of register 4292.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Command: Password

 

Modbus address: 4292

Length: 4

 

Type: 16

Write

 

Parameter:

 

Operator level

Pass number

 

 

 

 

 

 

 

Format:

 

Hex

 

Decimal

 

 

 

 

 

 

 

 

Value:

 

0x30

 

12345678

 

 

 

 

 

 

 

Figure 2.2.2.2: Example to set the Password of operator level S (code 0x30) to 12345678.

624300/01

page 6 / 67

Page 6
Image 6
Hamilton Electronics EPHUM011 manual Changing Passwords for Operator Level, 0x0C 18111978, 0x0B 18111978, 0x30 12345678