pH ARC Sensors Modbus RTU Programmer’s Manual (EPHUM011)

2.7.4.2.1Product calibration: Initial measurement

Calibration status after initial measurement command under conditions outside the valid calibration range for CP6 (defined in register 5312):

Command: Calibration status CP6

Modbus address: 5318

Length: 6

Type: 3

Read

Parameter:

Status

Unit

Value

 

 

 

Format:

Hex

Hex

Float

 

 

 

Value:

0x01000000

0x00001000

4.01

 

 

 

Figure 2.7.4.2.1.1: Example to read the calibration status of CP6 after having performed an initial measurement at CP6 under measurement conditions outside the calibration range for CP6. The status says: “CP6: out of calibration range”(0x01000000). The last successful calibration has been performed at pH 4.01.

The initial measurement in this case was not successful. The sensor is still running on its prior standard calibration.

Calibration status after successful initial measurement:

Command: Calibration status CP6

Modbus address: 5318

Length: 6

Type: 3

Read

Parameter:

Status

Unit

Value

 

 

 

Format:

Hex

Hex

Float

 

 

 

Value:

0x08000000

0x00001000

4.01

 

 

 

Figure 2.7.4.2.1.2: Example to read the calibration status of CP6 after having performed an initial measurement at CP6 under correct measurement conditions.

The status says: “CP6: initial measurement”(0x08000000). The last successful calibration has been performed at pH 4.01.

The initial measurement in this case was successful.

The sensor is still running on its prior standard calibration until a valid calibration value has been assigned to this initial measurement values.

2.7.4.2.2Product calibration: Assignment

Calibration status after invalid assignment:

Command: Calibration status CP6

Modbus address: 5318

Length: 6

Type: 3

Read

Parameter:

Status

Unit

Value

 

 

 

Format:

Hex

Hex

Float

 

 

 

Value:

0x0A000000

0x00001000

4.01

 

 

 

Figure 2.7.4.2.2.1: Example to read the calibration status of CP6 after having performed a valid initial measurement at CP6 and an invalid assignment.

The status says: “CP6: out of range”(0x02000000) and “CP6: initial measurement” (0x08000000). The last successful calibration has been performed at pH 4.01

The initial measurement in this case is still valid and available for further assignment of a product calibration value. The here performed assignment was not successful. The sensor remains running on its prior standard calibration.

Calibration status after valid assignment:

Command: Calibration status CP6

Modbus address: 5318

Length: 6

Type: 3

Read

Parameter:

Status

Unit

Value

 

 

 

Format:

Hex

Hex

Float

 

 

 

Value:

0x14000000

0x00001000

4.5

 

 

 

Figure 2.7.4.2.2.2: Example to read the calibration status of CP6 after having performed an initial measurement at CP6 and a valid assignment to pH 4.5.

The status says: “CP6: active”(0x04000000) and “CP6: assigned”(0x10000000). The last successful calibration corresponding to the here performed assignment has been performed at pH 4.5.

624300/01

page 46 / 67

Page 46
Image 46
Hamilton Electronics EPHUM011 0x01000000 0x00001000, 0x08000000 0x00001000, 0x0A000000 0x00001000, 0x14000000 0x00001000