pH ARC Sensors Modbus RTU Programmer’s Manual (EPHUM011)

2.7.7 Currently active Calibration Parameters part 3

In register 5520, 5528 and 5560 the pH value of the used calibration standards, the electrical potential of the pH sensor and the temperature upon calibration are stored.

Start

Number

Reg1 /

Reg3 /

Reg5 /

Reg7 /

Modbus

Read

Write

register

of

Reg2

Reg4

Reg6

Reg8

function

access

access

 

registers

 

 

 

 

code

 

 

5520

8

pH value of

Electrical

Electrical

free

3, 4

A/S

none

 

 

calibration

potential at

potential at

 

 

 

 

 

 

standard at

CP1

CP1

 

 

 

 

 

 

CP1 [pH]

[mV]

[mV]

 

 

 

 

5528

8

pH value of

Electrical

Electrical

free

3, 4

A/S

none

 

 

calibration

potential at

potential at

 

 

 

 

 

 

standard at

CP2

CP2

 

 

 

 

 

 

CP2 [pH]

[mV]

[mV]

 

 

 

 

5560

8

pH value of

Electrical

Electrical

free

3, 4

A/S

none

 

 

product at

potential at

potential at

 

 

 

 

 

 

CP6 [pH]

CP6 [mV]

CP6 [mV]

 

 

 

 

Figure 2.7.7.1: Definition of register 5520, 5528 and 5560.

Command: Act calibration CP1

Modbus address: 5520

 

Length: 8

Type: 3

Read

Parameter:

pH CP1 [pH]

Voltage CP1 [mV]

Temp

CP1 [K]

free

 

Format:

Float

Float

Float

 

Float

 

Value:

4.003401

179.927

297.1378

0

 

Figure 2.7.7.2:

Example to read the

actual calibration values

of CP1.

 

 

 

 

 

 

 

 

 

Command: Act calibration CP2

Modbus address: 5528

 

Length: 8

Type: 3

Read

Parameter:

pH CP2 [pH]

Voltage CP2 [mV]

Temp

CP2 [K]

free

 

Format:

Float

Float

Float

 

Float

 

Value:

7.006804

3.099747

296.6901

0

 

Figure 2.7.7.3: Example to read the actual calibration values of CP2.

Command: Act calibration CP6

Modbus address: 5560

Length: 8

Type: 3

Read

Parameter:

pH CP6 [pH]

Voltage CP6 [mV]

Temp

CP6 [K]

free

 

Format:

Float

Float

Float

 

Float

 

Value:

7.1

5.10469

298.3302

0

 

Figure 2.7.7.4: Example to read the actual calibration values of CP6.

2.7.8 Currently active Calibration Parameters part 4

For standard calibration (CP1 / CP2) register 5448 documents offset at pH7 and slope:

Start

Number

Reg1 / Reg2

 

Reg3 / Reg4

Reg5 / Reg6

 

Modbus

Read

 

Write

register

of

(Float)

 

(Float)

(Float)

 

 

function

access

 

access

 

registers

 

 

 

 

 

 

 

code

 

 

 

 

5448

6

 

Offset at

 

Slope (25 °C)

Reference

 

3, 4

 

U/A/S

 

none

 

 

 

pH 7 [mV]

 

[mV/pH]

temperature [K]

 

 

 

 

 

 

 

Figure 2.7.8.1: Definition of register 5448.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Command: Calculated cali values

 

Modbus address: 5448

Length: 6

 

Type: 3

Read

 

Parameter:

Offset at pH 7

Slope

 

Ref temp

 

 

 

 

 

 

 

 

[mV]

 

[mV/pH]

 

[K]

 

 

 

 

 

 

 

 

Format:

 

Float

 

Float

 

Float

 

 

 

 

 

 

 

 

Value:

 

3.607782

-59.47631

 

298.15

 

 

 

 

 

 

 

 

Figure 2.7.8.2: Example to read register 5448: offset at pH 7 is 3.6 mV; slope is -59.5 mV/pH; reference temperature is 298.15 K (=25°C)