pH ARC Sensors Modbus RTU Programmer’s Manual (EPHUM011)
The here performed assignment was successful. The sensor is running using a valid product calibration.
2.7.4.2.3Product calibration: Cancel
Calibration status after cancelling an active product calibration:
Command: Calibration status CP6 | Modbus address: 5318 | Length: 6 | Type: 3 | Read | ||
Parameter: | Status | Unit | Value |
|
|
|
Format: | Hex | Hex | Float |
|
|
|
Value: | 0x00000000 | 0x00001000 | 4.5 |
|
|
|
Figure 2.7.4.2.3.1: Example to read the calibration status of CP6 after having performed a cancel command at CP6.
The status reports no messages. The last successful calibration at CP6 has been performed at pH 4.5.
The sensor is running on a valid standard calibration and no product calibration is stored.
2.7.4.2.4Product calibration: Restore standard calibration
Calibration status after restoring a standard calibration from an active product calibration:
Command: Calibration status CP6 | Modbus address: 5318 | Length: 6 | Type: 3 | Read | ||
Parameter: | Status | Unit | Value |
|
|
|
Format: | Hex | Hex | Float |
|
|
|
Value: | 0x10000000 | 0x00001000 | 4.5 |
|
|
|
Figure 2.7.4.2.4.1: Example to read the calibration status of CP6 after having restored the standard calibration from an active product calibration (CP6).
The status says: “CP6 assigned”(0x10000000). The last successful calibration at CP6 has been performed at pH 4.5.
The sensor is running on a valid standard calibration but a valid product calibration is still available in the sensor.
2.7.4.2.5Product calibration: Restore product calibration
Calibration status after restoring an available product calibration from an active standard calibration:
Command: Calibration status CP6 | Modbus address: 5318 | Length: 6 | Type: 3 | Read | ||
Parameter: | Status | Unit | Value |
|
|
|
Format: | Hex | Hex | Float |
|
|
|
Value: | 0x14000000 | 0x00001000 | 4.5 |
|
|
|
Figure 2.7.4.2.5.1: Example to read the calibration status of CP6 after having restored an available product calibration (CP6) from an active standard calibration (CP1 / CP2).
The status says: “CP6: active”(0x04000000) and “CP6: assigned”(0x10000000). The last successful calibration corresponding to the here performed assignment has been performed at pH 4.5.
The sensor is running on a valid product calibration again.
624300/01 | page 47 / 67 |