pH ARC Sensors Modbus RTU Programmer’s Manual (EPHUM011)

2.4Configuration of the Analog Interfaces

2.4.1 Available Analog Interfaces

A pH ARC Sensor has two individual physical analog interfaces that have identical functionalities, but can be configured independently from each other.

Analog Output Interface 1 (AO1)

Analog Output Interface 2 (AO2)

The number of analog interfaces is defined in register 4320.

Start

Number of

Reg1 / Reg2

Modbus

Read

Write

register

registers

 

function code

access

access

4320

2

Available analog interfaces

3, 4

U/A/S

none

Figure 2.4.1.1: Definition of register 4320.

Command: Avail analog interfaces

Modbus address: 4320

Length: 2

Type: 3

Read

Parameter:

Available analog

 

 

 

 

 

 

 

interfaces

 

 

 

 

 

 

Format:

Hex

 

 

 

 

 

 

Value:

0x03

 

 

 

 

 

 

Figure 2.4.1.2: Example to read the available analog interfaces. The answer is “0x03”meaning that there exists an Analog Interface 1 (AO1) and an Analog Interface 2 (AO2).

2.4.2 Available Analog Interface Modes

With register 4322, the available analog interface modes for AO1 and AO2 are defined

Start

 

Number of

 

Reg1 /

 

Reg3 /

Reg5 /

Reg7 /

 

Modbus

Read

 

Write

register

 

registers

 

Reg2

 

Reg4

Reg6

Reg8

 

function

access

 

access

 

 

 

 

 

 

 

 

 

 

 

code

 

 

 

4322

8

 

Available

 

Available

reserved

reserved

 

3,4

U/A/S

 

none

 

 

 

 

Analog

 

Analog

 

 

 

 

 

 

 

 

 

 

 

Interface

 

Interface

 

 

 

 

 

 

 

 

 

 

 

Modes for

Modes

 

 

 

 

 

 

 

 

 

 

 

AO1

 

for AO2

 

 

 

 

 

 

 

Figure 2.4.2.1: Definition of register 4322. It defines the analog interface modes available for AO1 and

AO2. The analog interface modes are described in Figure 2.4.2.2.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Code

 

Analog Interface

 

Description

 

 

 

 

 

 

 

(Hex)

 

Mode

 

 

 

 

 

 

 

 

 

 

 

0x00

 

4-20 mA inactive

 

Analog interface deactivated

 

 

 

 

 

 

0x01

 

4-20 mA fixed

 

 

Set to a constant output value for current loop testing

 

 

 

0x02

 

4-20 mA linear

 

Linear output of measurement (PMC1 / 6)

 

 

 

 

0x04

 

4-20 mA bilinear

 

Bilinear output of measurement (PMC1 / 6)

 

 

 

 

Figure 2.4.2.2: Definition of the analog interface modes, valid for both AO1 and AO2.

Command: Analog Interface Modes

Modbus address: 4322

Length: 8

Type: 3

Read

Parameter:

Available Analog

Available Analog

reserved

 

reserved

 

 

Interface Modes for

Interface Modes for

 

 

 

 

 

AO1

AO2

 

 

 

 

Format:

Hex

Hex

Hex

 

Hex

 

Value:

0x07

0x07

0x0

 

0x0

 

Figure 2.4.2.3: Example to read register 4322: all modes defined in figure 2.4.2.2 are available for both AO1 and AO2.

624300/01

page 9 / 67

Page 9
Image 9
Hamilton Electronics EPHUM011 manual Configuration of the Analog Interfaces, Available Analog Interfaces, 0x07