pH ARC Sensors Modbus RTU Programmer’s Manual (EPHUM011)

2.8.3 Warnings

A “Warning”is a notification message which still allows further functioning of the system. This message alerts the operator of a possible problem that could lead to uncertain results.

2.8.3.1Currently Active Warnings

The currently active warnings are stored in register 4736.

Start

Number of

Reg1 /

Reg3 /

Reg5 /

Reg7 /

Modbus

Read

Write

register

registers

Reg2

Reg4

Reg6

Reg8

function

access

access

 

 

 

 

 

 

code

 

 

4736

8

Active

Active

Active

Active

3, 4

U/A/S

none

 

 

warning

warning

warning

warning

 

 

 

 

 

measure-

calibration

interface

hardware

 

 

 

 

 

ment

(bitwise

(bitwise

(bitwise

 

 

 

 

 

(bitwise

defined)

defined)

defined)

 

 

 

 

 

defined)

 

 

 

 

 

 

Figure 2.8.3.1.1: Definition of register 4736 (see chapter 2.8.3.3)

Command: Active warning

 

Modbus address: 4736

Length: 8

Type: 3

Read

Parameter:

W Measurement

W

Calibration

W Interface

W Hardware

 

Format:

Hex

Hex

Hex

 

Hex

 

Value:

0x00

0x00

0x00

 

0x00

 

Figure 2.8.3.1.2: Example to read the currently active warnings.

2.8.3.2History of Warnings

The history of warnings is not implemented in pH ARC Sensors.

2.8.3.3Definition of Warnings

Bit #

Hex

Description

 

 

not available

Figure 2.8.3.3.1: Definition of warnings “measurement”. None is defined.

Bit #

Hex

Description

0 (LSB)

0x0001

PMC1 (pH) calibration recommended

1

0x0002

PMC1 (pH) last calibration not successful

Figure 2.8.3.3.2: Definition of warnings “calibration”.

 

 

 

Bit #

Hex

Description

 

 

not available

Figure 2.8.3.3.3: Definition of warnings “interface”. None is defined.

Bit #

Hex

Description

 

 

not available

Figure 2.8.3.3.4: Definition of warnings “hardware”. None is defined.