Hamilton Electronics EPHUM011 manual Reading the Sensor’s Quality Indicator, 120 130

Models: EPHUM011

1 67
Download 67 pages 49.01 Kb
Page 63
Image 63

pH ARC Sensors Modbus RTU Programmer’s Manual (EPHUM011)

Command: SIP definition

 

Modbus address: 4988

Length: 8

Type: 16

Write

Parameter:

T min [°C]

T

max [°C]

Time min [min]

Empty

 

Format:

Float

Float

Float

 

Float

 

Value:

120

130

30

 

0

 

Figure 2.8.5.3:

Example to write the

SIP definitions.

 

 

 

 

 

 

 

 

 

 

Command: SIP definition

 

Modbus address: 4988

Length: 8

Type: 3

Read

Parameter:

T min [°C]

T

max [°C]

Time min [min]

Empty

 

Format:

Float

Float

Float

 

Float

 

Value:

120

130

30

 

0

 

Figure 2.8.5.4:

Example to read the

SIP definitions.

 

 

 

 

 

 

 

 

 

 

Command: CIP definition

 

Modbus address: 4996

Length: 8

Type: 16

Write

Parameter:

T min [°C]

T

max [°C]

Time min [min]

Empty

 

Format:

Float

Float

Float

 

Float

 

Value:

80

100

30

 

0

 

Figure 2.8.5.5:

Example to write the

CIP definitions.

 

 

 

 

 

 

 

 

 

 

Command: CIP definition

 

Modbus address: 4996

Length: 8

Type: 3

Read

Parameter:

T min [°C]

T

max [°C]

Time min [min]

Empty

 

Format:

Float

Float

Float

 

Float

 

Value:

80

100

30

 

0

 

Figure 2.8.5.6: Example to read the CIP definitions.

2.8.6 Reading the Sensor’s Quality Indicator

In register 4872 the sensor’s quality indicator (0-100%) is given.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Start

Number of

 

Reg1 / Reg2

 

Modbus

Read

 

Write

 

 

 

register

registers

 

(Float)

 

function

access

access

 

 

 

 

 

 

 

 

 

code

 

 

 

 

 

 

4872

2

 

 

Quality [%]

 

3, 4

U/A/S

none

 

 

 

Figure 2.8.6.1: Definition of register 4872.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Command: Quality indicator

Modbus address: 4872

Length: 2

 

Type: 3

Read

Parameter:

Quality [%]

 

 

 

 

 

 

 

 

 

Format:

 

Float

 

 

 

 

 

 

 

 

 

Value:

 

100

 

 

 

 

 

 

 

 

 

 

Figure 2.8.6.2: Example to read the sensor’s quality indicator.

The sensor’s quality indicator is influenced by:

Offset a pH 7

Slope

Errors

624300/01

page 63 / 67

Page 63
Image 63
Hamilton Electronics EPHUM011 manual Reading the Sensor’s Quality Indicator, 120 130