pH ARC Sensors Modbus RTU Programmer’s Manual (EPHUM011)
2.3Configuration of the serial RS485 Interface
Factory settings of the RS485:
Parity is none, 1 start bit, 8 data bits, 2 stop bits (in total: 11 bits).
2.3.1Device Address
2.3.1.1Reading and Writing the Device Address
The sensor specific device address can be read and written in register 4096.
Start |
| Number of |
| Reg1 / Reg2 | Modbus | Read | Write |
|
| ||||
register |
| registers |
|
|
| function code | access | access |
|
| |||
4096 |
| 2 |
|
| device address | 3, 4, 16 |
| U/A/S | S |
|
| ||
Figure 2.3.1.1.1: Definition of register 4096. |
|
|
|
|
|
|
| ||||||
|
|
|
|
|
|
|
| ||||||
Command: Com address |
| Modbus address: 4096 | Length: 2 | Type: 3 | Read | ||||||||
Parameter: |
| Modbus address |
|
|
|
|
|
|
|
| |||
Format: |
| Decimal |
|
|
|
|
|
|
|
|
| ||
Value: |
| 1 |
|
|
|
|
|
|
|
|
|
|
Figure 2.3.1.1.2: Example to read the device address.
The device address can be set by S (Specialist), default value is 1.
Command: Com address | Modbus address: 4096 | Length: 2 | Type: 16 | Write | |||
Parameter: | Modbus address |
|
|
|
|
|
|
Format: | Decimal |
|
|
|
|
|
|
Value: | 3 |
|
|
|
|
|
|
Figure 2.3.1.1.3: Example to set the device address to 3.
2.3.1.2Reading the Device Address Limits
The device address limits can be read in register 4098.
Start |
| Number of | Reg1 / Reg2 |
| Reg3 / Reg4 | Modbus | Read | Write |
| ||||
register |
| registers |
|
|
|
|
| function code | access | access |
| ||
4098 |
| 4 |
| Min. device |
| Max. device | 3, 4 |
| U/A/S | none |
| ||
|
|
|
| address |
| address |
|
|
|
|
| ||
Figure 2.3.1.2.1: Definition of register 4098. |
|
|
|
|
| ||||||||
|
|
|
|
|
|
|
|
| |||||
Command: Com address limits | Modbus address: 4098 | Length: 4 |
| Type: 3 | Read | ||||||||
Parameter: |
| Min value |
| Max value |
|
|
|
|
|
| |||
Format: |
| Decimal |
| Decimal |
|
|
|
|
|
| |||
Value: |
| 1 |
| 32 |
|
|
|
|
|
|
|
|
Figure 2.3.1.2.2: Example to read the device address limits: Min = 1, Max = 32.
624300/01 | page 7 / 67 |