pH ARC Sensors Modbus RTU Programmer’s Manual (EPHUM011)

2.7.10Calibration Standards

2.7.10.1 Available Sets of Calibration Standards

pH ARC Sensors can store and operate six different sets of calibration standards each having max 12 calibration standards. Before calibration the operator has to select one defined set to operate with. Default setting is the HAMILTON set of calibration standards.

The operator can change the selection of sets at any time, without compromising prior calibration data at CP1 / CP2.

In register 9472 the available sets of calibration standards are defined.

Start

Number of

Reg1 / Reg2

Modbus

Read

Write

register

registers

(Bit, see Figure 2.7.10.1.2)

function code

access

access

9472

2

Available calibration standard sets

3, 4

U/A/S

none

Figure 2.7.10.1.1: Definition of register 9472.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Bit #

 

Hex value

Description

Definition

 

 

 

 

0 (LSB)

 

0x00000001

Set 1

HAMILTON

 

 

 

 

1

 

0x00000002

Set 2

MERCK TITRISOL

 

 

 

 

2

 

0x00000004

Set 3

DIN 19267

 

 

 

 

3

 

0x00000008

Set 4

NIST Standard

 

 

 

 

4

 

0x00000010

Set 5

METTLER-TOLEDO

 

 

 

 

5

 

0x00000020

Set 6

RADIOMETER

 

 

 

 

Figure 2.7.10.1.2: Definition of available sets of calibration standards.

 

 

 

 

 

 

 

 

 

 

 

Command: Available cali sets

 

Modbus address: 9472

Length: 2

 

Type: 3

Read

Parameter:

 

Calibration sets

 

 

 

 

 

 

 

Format:

 

Hex

 

 

 

 

 

 

 

Value:

 

0x0000003F

 

 

 

 

 

 

 

Figure 2.7.10.1.3: Example to read the available calibration sets: Set 1 (0x00000001) + set 2 (0x00000002) + set 3 (0x00000004) + set 4 (0x00000008) + set 5 (0x00000010) + set 6 (0x00000020), in total 0x0000003F.

In register 9474 the selected set of calibration standards is defined.

Start

Number of

Reg1 / Reg2

Modbus

Read

Write

 

register

registers

(Bit, see Figure 2.7.10.1.2)

function code

access

access

 

9474

2

 

Selected set of calibration standard

3, 4

U/A/S

S

 

Figure 2.7.10.1.4: Definition of register 9474. Only one bit can be set.

 

 

 

 

 

 

 

 

 

Command: Selected set of cal stand.

Modbus address: 9474

Length: 2

Type: 3

Read

Parameter:

Calibration set

 

 

 

 

 

Format:

 

Hex

 

 

 

 

 

 

Value:

 

0x00000001

 

 

 

 

 

Figure 2.7.10.1.5: Example to read the selected set of calibration standards. Set 1 (HAMILTON) (0x00000001) is active.

Command: Select set of cal stand.

Modbus address: 9474

Length: 2

Type: 16

Write

Parameter:

Calibration set

 

 

 

 

Format:

Hex

 

 

 

 

Value:

0x04

 

 

 

 

Figure 2.7.10.1.6: Example to set the calibration standard set to DIN 19267 (0x04).

624300/01

page 51 / 67

Page 51
Image 51
Hamilton Electronics EPHUM011 manual Available Sets of Calibration Standards, LSB 0x00000001 Set, 0x0000003F