pH ARC Sensors Modbus RTU Programmer’s Manual (EPHUM011)

Command: Info user

Modbus address: 1568

Length: 8

Type: 16

Write

Parameter:

Text

 

 

 

 

 

Format:

Character

 

 

 

 

 

Value:

Hello World

 

 

 

 

 

Figure 2.9.3.2:

Example to write 16

ASCII characters to register 1568 with operator A or S.

 

 

 

 

 

 

Command: Info user

Modbus address: 1568

Length: 8

Type: 3

Read

Parameter:

Text

 

 

 

 

 

Format:

Character

 

 

 

 

 

Value:

Hello World

 

 

 

 

 

Figure 2.9.3.3: Example to read the register 1568 (written in figure 2.9.1.6).

2.10 System Commands

2.10.1 Recall Sensor’s Factory Settings

Using register 8192 you can recall the sensor manufacturer values (interfaces, calibration data and passwords), except the SIP and CIP data which remain unchanged. By sending the recall value “732255”, all configuration values will be set to default.

Start

 

Number of

Reg1 / Reg2

Modbus

 

Read

Write

 

register

 

registers

 

 

 

function code

 

access

access

 

8192

 

2

 

Recall by value 732255

16

 

 

none

S

 

Figure 2.10.1.1: Definition of register 8192.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Command: Recall

 

Modbus address: 8192

Length: 2

 

Type: 16

Write

 

Parameter:

 

Recall

 

 

 

 

 

 

 

 

 

Format:

 

Decimal

 

 

 

 

 

 

 

 

 

Value:

 

732255

 

 

 

 

 

 

 

 

 

Figure 2.10.1.2: Example to write the restore command.

3

Abbreviations

AO

Analog Output Interface

CP

Calibration Point

ECS

Electrochemical Sensor Interface

PMC

Primary Measurement Channel

SMC

Secondary Measurement Channel

624300/01

page 66 / 67

Page 66
Image 66
Hamilton Electronics EPHUM011 manual System Commands, Recall Sensor’s Factory Settings, Hello World, 732255