Storing, Logging and Printing
Logging Results
Logging Results
Description Test Period Logging
If degradations in system performance can be observed at an early stage, then the appropriate remedial action can be taken to maximize circuit availability and avoid system crashes. Test period logging allows you to monitor the error performance of your circuit. At the end of the test period the selected results are logged. Results can be logged at regular intervals during the test period by selecting a LOGGING PERIOD of shorter duration than the test period. An instant summary of the results can be demanded by pressing PRINT NOW without affecting the test in progress.
Error Event Logging
Manual tracing of intermittent faults is time consuming. Error event logging allows you to carry out unattended long term monitoring of the circuit. Each occurrence of the selected error event is logged.
The results obtained during the test are retained in memory until they are overwritten by the next set of results. The results can be logged at any time during the test period and at the end of the test period. The results required are selected using OTHER LOGGING LOGGING SETUP CONTROL.
Any Alarm occurrence results in a timed and dated message being logged.
BER and Analysis results can be selected by the user.
Cumulative and Period versions of the results are calculated and can be selected by the user.
Period | The results obtained over a set period of time during |
| the test. The Period is defined by the LOGGING |
| PERIOD selection. |
Cumulative | The results obtained over the time elapsed since the |
| start of the test. |
The results can be logged to the following devices, selectable using
OTHER LOGGING LOGGING SETUP DEVICE :
107