Selecting and Using "Other" Features

Running Self Test

Running Self Test

Description Before using the HP 37718A to make measurements, you can run Self Test ALL TESTS to ascertain the integrity of the HP 37718A. These tests take between at least 1 hour to complete depending on the options fitted.

Alternatively you can run Confidence Tests which only takes 2 to 3 minutes to complete. This is not a full verification but performs BER measurements with internal and external loopbacks fitted.

HOW TO:

Run ALL TESTS

 

1

Choose TEST TYPE

 

on the OTHER

 

 

ALL TESTS

SELF TEST

 

2

Insert a formatted disk into the instrument disk drive.

display.

3Make the loopback connections listed below:

Connect Transmit module 75Ω OUT to Receive module 75Ω IN Connect Transmit module 100/120Ω OUT to Receive module 100/ 120Ω IN

Connect Transmit module 75Ω MUX to Receive module 75Ω DEMUX Connect Transmit module 100/120Ω MUX to Receive module 100/ 120Ω DEMUX

Connect Multirate Analyser IN to OUT

Connect Optical OUT to Optical IN via a 15 dB attenuator.

163

Page 173
Image 173
HP BER 718 37718A manual Running Self Test, Run ALL Tests