Storing, Logging and Printing

Logging Jitter Tolerance Results

Logging Jitter Tolerance Results

Description The jitter auto tolerance feature provides jitter tolerance measurements within the relevant ITU-T mask, G.823 for DSn, G.958, G.825 and G.253 for SONET.

You can log the jitter auto tolerance results to a printer for record

keeping purposes. The GRAPH version or the TEXT version of the jitter

tolerance results can be logged to an external printer.

HOW TO:

1 If logging to a Parallel Port (Centronics) printer, connect the printer to

 

the PARALLEL port. See "Connecting an HP 850C DeskJet Printer to

 

a Parallel Port " page 121.

2If a non-HP printer is connected choose ALT. PRINTER .

3Choose 80 character column width ( NORMAL ) or 40 character column width ( COMPRESS ) according to the capabilities of your printer.

4If logging to an HP-IB printer, connect an HP-IB printer to the HPIB port. See "Logging Results to HP-IB Printer " page 117.

5If logging to an RS-232-C printer, connect an RS-232-C printer to the RS232 port. See "Logging Results to RS-232-C Printer " page 119.

6If a non HP printer is connected, choose

ALT. PRINTER

.

112

Page 122
Image 122
HP BER 718 37718A manual Logging Jitter Tolerance Results