Storing, Logging and Printing
Logging Results
•Optional Internal printer fitted into the instrument front cover (Option 602)
•External
•External
•External Parallel Port printer (option 601)
•Disk Drive
HOW TO: | 1 | Choose LOGGING [ON] - enables the logging of results and alarms. |
| 2 | Choose LOGGING PERIOD - determines how regularly the results |
|
| and alarms are logged. |
|
| USER PROGRAM provides a choice of 10 minutes to 99 hours. |
| 3 | Choose RESULT LOGGED - allows you to log all results to or choose |
|
| only those results you require. |
| 4 | Choose WHEN - allows you to choose to only log when the error count |
|
| for the logging period is greater than 0. If the error count is 0 then the |
|
| message NO BIT ERRORS is displayed. |
| 5 | Choose CONTENT - allows you a choice of error results to be logged. |
|
| Error Results, Analysis or Error and Analysis (ER & ANAL) |
|
| and |
|
| Period, Cumulative or Period and Cumulative (PER & CUMUL). |
| 6 | If LOG ERROR SECONDS [ON] is chosen a timed and dated message |
|
| is logged each time an error second occurs (excessive occurrences of |
|
| error seconds during the logging period will result in heavy use of |
108