SECTION 5—FRONT RIGGINGS

Installing/Adjusting the Adjustable Angle Flip-up Footplates

WARNING

When determining the angle of the footplates, make sure the rear of the footplates do not interfere with the movement of the front casters.

NOTE: For this procedure, refer to

FIGURE 5.7.

Installing Adjustable Angle Flip-up Footplate

1.Slide the half clamp over the footplate hinge.

2.Loosely tighten the two flat screws that secure the footplate to the half clamp.

3.Adjust the footplates to the necessary angle and depth for the user. Refer to the following sections of this procedure.

Articulating

Flat

90o Footrest

Footplate

Screws

Support

Nylon

Adjustment

Screw

Half

Clamp Footplate

Hinge

Washers

Half

Locknuts

Clamp Footplate

Hinge

 

FIGURE 5.7 Installing Adjustable Angle

Flip-up Footplate

Adjusting Depth on Adjustable Angle Flip-Up

NOTE: For this procedure, refer to FIGURE 5.8 on page 35.

NOTE: Observe the angle of the articulating footplate for reinstallation.

1.Remove the two flat screws, washers and locknuts that secure articulating footplate to the footplate hinge.

2.Move the articulating footplate to one of four mounting positions.

NOTE: If desired depth is still not obtained, rotate the half clamp on the footplate hinge 180o. 3. Retighten the two flat screws, washers and locknuts.

NOTE: The settings for positioning the articulating footplates on the half‐clamps may vary for each footplate.

Adjusting Angle on Adjustable Angle Flip-up Footplate

NOTE: For this procedure, refer to FIGURE 5.8 on page 35.

1.Loosen, but do not remove the two flat screws and locknuts that secure the footplate to the footplate hinge.

2.Position the adjustable angle flip‐up footplate to the desired angle.

3.Retighten the two flat screws and locknuts.

Orbit

34

Part No. 1073955

Page 34
Image 34
Invacare 1073955 manual Installing Adjustable Angle Flip-up Footplate, Adjusting Depth on Adjustable Angle Flip-Up