SECTION 6—ARMS

Replacing/Repositioning the Locking Mechanism in the Cantilever Arm

NOTE: For this procedure, refer to FIGURE 6.10 on page 50.

CAUTION

The locking mechanism is spring loaded. Place your free hand over the locking mechanism to prevent parts from springing out of the armrest.

1.Move the cantilever arm up and out of the way.

2.Unthread the actuator from the locking mechanism and remove.

3.Slowly let the locking mechanism and spring slide out of the cantilever arm.

NOTE: Inspect the spring for wear and damage. Replace if necessary.

4.Slide the new locking mechanism and spring into the cantilever arm.

5.Position the angled portion of the locking mechanism in one of two ways:

Angled Portion Facing Up ‐ Arm will flip up

Angled Portion Facing Down ‐ Arm will flip down

6.Use Loctite 242 and securely tighten the actuator into the locking mechanism.

7.To lock the cantilever arm, push down (or pull up) until there is an audible click.

8.Pull up (or push down) on the cantilever arm to make sure it is locked in place.

Angled Portion of

Locking Mechanism

Locking

 

Spring

 

 

 

Mechanism

 

 

Cantilever Arm

Actuator

(Apply Loctite 242)

FIGURE 6.10 Replacing/Repositioning the Locking Mechanism in the Cantilever Arm

Orbit

50

Part No. 1073955

Page 50
Image 50
Invacare 1073955 manual Angled Portion Locking Mechanism