SECTION 5—FRONT RIGGINGS

Footrest Support

Front View of

Footplate and

 

 

Footrest

Footplate

Support

 

FIGURE 5.8 Adjusting Depth on Adjustable Angle Flip-Up- Adjusting Angle on Adjustable

Angle Flip-up Footplate

Adjusting Perpendicular and/or Inversion/Eversion Adjustable Angle Flip-up Footplate

NOTE: For this procedure, refer to FIGURE 5.9.

NOTE: It is not necessary to remove the footplate to perform this adjustment.

1.Insert a flathead screwdriver through the half clamp on the adjustable angle footplate.

2.Slowly turn the nylon adjustment screw in or out until the adjustable angle footplate is perpendicular to the footrest assembly or the desired inversion or eversion is obtained.

Footrest

Side View Of

Footplate and

Support

Footrest

 

Footplate

 

FIGURE 5.9 Adjusting Perpendicular and/or Inversion/Eversion Adjustable Angle Flip-up Footplate

Bilateral Contracture Platform Procedures

WARNING

The footrest assembly MUST be at least 1-3/4-inches above the ground/floor to avoid hitting protruding objects when using this wheelchair.

NOTE: For this procedure, refer to FIGURE 5.10 on page 36.

Installing Bilateral Contracture Platform

NOTE: If a seating system is being used on the chair, refer to the seating system Ownerʹs Manual for installation and removal of the seating system.

1.Remove the existing seating system from the wheelchair, if so equipped.

2.Remove existing footrest from the wheelchair. Refer to Installing/Removing Front Riggings on page 30.

3.Remove the telescoping tubes from the wheelchair. Refer to Installing/Removing/Adjusting the Telescoping Front Frame Tubes on page 38.

Part No. 1073955