SECTION 5—FRONT RIGGINGS

Angle

Locknuts

Mounting

Adjustment

Plate

Screw

Hole #

 

4

Upper

3

Support of

2

Bilateral

1

Contracture

 

Angle

Platform

NOTE: One mounting screw removed for clarity.

FIGURE 5.11 Adjusting Height of Bilateral Contracture Platform

Adjusting Bilateral Contracture Platform Angle

1.Remove the two mounting screws and locknuts that secure the upper support of the bilateral contracture platform to the angle adjustment plates.

2.Reposition the bilateral contracture platform to one of four positions depending on the angle needed:

HOLE #

1

2

3

4

 

 

 

 

 

ANGLE

60°

75°

90°

118°

 

 

 

 

 

NOTE: Holes numbered from the bottom of the wheelchair towards the top of the wheelchair for reference only. (There are no numbers on the wheelchair frame.)

3.Reinstall the two mounting screws and locknuts that secure the upper support of the bilateral contracture platform to the angle adjustment plates and securely tighten.

NOTE: If the angle needed is still not obtainable, refer to Adjusting Angle on Adjustable Angle Flip‐up Footplate on page 34.

Adjusting Bilateral Contracture Platform Footplate Depth

NOTE: Refer to Adjusting Depth on Adjustable Angle Flip‐Up on page 34.

Replacing Sector Block

NOTE: For this procedure, refer to FIGURE 5.12 on page 38.

1.Remove the mounting screw and washer that secure the existing sector block to the wheelchair frame.

2.With locking pin facing up, secure the new sector block to the wheelchair frame with the existing mounting screw and washer. Use Loctite 242® and securely tighten.

Part No. 1073955

37

Orbit

Page 37
Image 37
Invacare 1073955 manual Replacing Sector Block, Adjusting Bilateral Contracture Platform Angle