A-61408 September 1995 C-11
The following action(s) should be taken:
• Verify the position of the bar code placed on the Imaging
Guide.
• Clean the Imaging Guides if required.
NOTE: Press the Stop or End key to return to the Function
65 menu.
(C) – (1 )
W/N = XX. XX : 1
NEW = . XXXX
This display gives the first page of the results of the bar code
evaluation. The results consists of three pages. Each item will
be described later on.
18. Press the Enter key to proceed to the next page of the
results.
(C) – (2)
R M I N = X X X %
C o n t r a s t = X X X %
This display gives the second page of the results.
19. Press the Enter key to proceed to the next page of the
results.
(C) – (3)
M O D = X X X %D E F = X X X % G r a d e = XThis display gives the third (final) page of the results. The last
item displayed is an overall grade which is based upon all six
field values. The fields which appeared in the three pages are
of two types: dimensional fields and reflectance fields. Each
of the reflectance field values are converted to a letter grade
(where A is the highest and F is the lowest). The lowest letter
grade assigned to any one of the reflectance fields is the
overall grade assigned to the bar code.
20. Upon completion of the test or from any of the displays
mentioned above, press the Stop key in order to return to the
menu page of Function 65; the following message will be
displayed to remind the operator to remove the bar code from
the Imaging Guide:
R e m o v e D o c , C l o s eD o o r , P r e s s E n t e rNOTE: One of the access doors must be opened and closed
and then the Enter key pressed to return to the
Function 65 menu, or one of the access doors opened
and closed and then the Stop key pressed to exit the
function.