Measurement Specialties 9116 user manual Span Calibration Valve Control

Models: 9116

1 143
Download 143 pages 39.58 Kb
Page 104
Image 104

Pressure Systems, Inc.

Model 9116 User’s Manual

 

 

For Model 9116, with integral DH200 transducers, internal manifolds and valves allow a Span adjustment pneumatic calibration to be accomplished easily and automatically.

It is recommended that a Span adjustment be performed whenever new transducers are installed in the instrument. In such cases, the new gain coefficients that result should always be stored back into the transducer’s non-volatile memory. In other cases, the user’s application may not require periodic span adjustment since the other factory-determined pressure/temperature coefficients (stored permanently inside each transducer) are extremely stable. Only an occasional Re-zero adjustment may be all that is necessary.

When instructed to execute a Calculate and Set Gains (‘Z’) command, the module will perform the Span adjustment calibration, and then update the gain coefficients in its memory. It will subsequently use the newly calculated gain terms for subsequent engineering-unit calculations until power is lost to the module.

Note

When using the Calculate and Set Gain (‘Z’) command, only the local variables in the module’s volatile main memory (RAM) are changed.

4.3.1.Span Calibration Valve Control

Before executing a Span adjustment (Calculate and Set Gains (‘Z’) command), the Model 9116 modules should have their calibration manifold valve placed in the proper position. For single pressure range units the CAL position should be used since the span calibration pressure can be applied between the CAL and CAL REF ports. Since the module will not attempt to shift this valve automatically, as it does for Re-zero adjustment, it should manually be placed in the desired position with the Set Operating Options (‘w’) command (option indexes = 0C and 12).

When span calibrating Model 9116 modules with multiple ranges installed, the CAL port may be used to apply pressure to all transducers only if the specified proof pressure is not exceeded on any channel. Refer to Calculate and Set Gains (‘Z’) command to specify the channels to be affected by the command in a multi-range unit. If the application of a specific span pressure exceeds the proof pressure rating of any other transducer contained within the same scanner, the calibration pressures must be applied to the RUN side pneumatic input ports. Since the calibration command (‘Z’) has a channel selection bit map parameter allowing it to calibrate only the desired pressure channels, the RUN port is a viable option for supplying the calibration pressures.

Page 95

www.PressureSystems.com

Page 104
Image 104
Measurement Specialties 9116 user manual Span Calibration Valve Control