Pressure Systems, Inc.

Model 9116 User’s Manual

 

 

The optional ipaddr parameter is normally unspecified, causing it to default to use the IP address of the current TCP/IP connection. That way the host need not have to be aware of its own IP address. This parameter is provided in case a special host has multiple network interfaces and wants to use more than one. When used, ipaddr requires four dotted numeric fields (d.d.d.d). Each d is a 1-3 digit decimal number in the range 0-255. The ipaddr = 255.255.255.255 is best avoided, unless the UDP datagrams of streams are to be broadcast to all network nodes.

Example:

Configure all streams to be delivered via UDP/IP protocol. Host expects the UDP datagrams to arrive via port 7500. The IP Address of the current TCP/IP connection is also used to send each UDP datagram.

“c 06 0 1 7500” Read response:

“A”

Configure all streams to be delivered via the default TCP/IP protocol.

“c 06 0 0” Read response:

“A”

Page 62