Measurement Specialties 9116 user manual Calibration Manifold Piston O-Ring Replacement

Models: 9116

1 143
Download 143 pages 39.58 Kb
Page 124
Image 124

Pressure Systems, Inc.

Model 9116 User’s Manual

 

 

5.1.6.4Calibration Manifold Piston O-Ring Replacement

Following is a step-by-step procedure to replace Calibration Manifold O-rings in a Model 9116 Intelligent Pressure Scanner. There are eight (8) pistons, each with an O-ring, inside the calibration valve housing; one (1) on each end of the housing, and three (3) on each side of the calibration valve itself.

(1)Disassemble the module as described in Section 5.1.2.

(2)Remove the PC-327 Analog board as described in Section 5.1.3.1. Lay the circuit board aside on an anti-static surface.

(3)Remove the tubing plate as described in Section 5.1.6.3.

(4)Using your index finger, shift the calibration manifold back and forth several times to loosen its connection with the adapter plate O-rings. Carefully lift the calibration valve housing with one hand and turn it over, letting the calibration manifold fall into the free hand. It is imperative that the calibration manifold does not fall on a hard surface since scratches on the manifold can result in pneumatic leaks.

(5)To remove the six (6) pistons from their slots on the manifold valve, hold the valve in one hand and apply air pressure of approximately 30 psi (200 kPa) to the C3/C4 input passages on the valve. The passages (0.043" diameter holes) are located on the side of the valve next to the tubing plate, one hole on each side of the valve. Pressure on one side will release three (3) pistons, and pressure on the other side will release the other three (3). If the pistons stick, apply a slightly higher pressure. Place your free hand over the calibration valve housing to prevent the pistons from coming out of the housing. Thoroughly clean the pistons with a fast evaporating cleaning fluid that leaves little or no residue (e.g., acetone, alcohol, Freon®, etc.) and dry with supply air. Replace the piston O-rings after lightly lubricating the rings with Krytox® fluorinated grease. Replace the pistons in their cavity by placing the O-ring side of the piston in first and then pressing the piston completely into its cavity with one finger.

(6)To remove the two (2) C1/C2 pistons from their slots on either end of the manifold valve housing, hold the manifold valve housing in one hand, and apply approximately 30 psi (200 kPa) to the two bulge tubes, one on either end of the housing end-plate. This will result in the pistons being forced out of their cavity. If the pistons stick, apply a higher pressure. Place your free hand over the calibration valve housing to prevent the pistons from coming out of the housing. Clean the pistons, lubricate and replace the O-rings, and replace the pistons into their cavities as described in (5) above.

(7)Thoroughly clean the calibration manifold with a fast-evaporating cleaning fluid that leaves little or no residue (e.g., alcohol, acetone, Freon®, etc.). Replace the calibration manifold into the housing, making sure that the guiding pin fits into the slot of the manifold housing.

(8)Replace the tubing plate as described in Section 5.1.6.3.

(9)Replace the PC-327 Analog board as described in Section 5.1.3.1 and reassemble the module.

(10)Test your scanner to ensure proper operation.

Page 115

www.PressureSystems.com

Page 124
Image 124
Measurement Specialties 9116 user manual Calibration Manifold Piston O-Ring Replacement