Pressure Systems, Inc.

Model 9116 User’s Manual

 

 

3.1.3Responses

Four (4) types of responses can be returned from a Model 9116 Intelligent Pressure Scanner module:

an Error response,

an Acknowledge response,

an Acknowledge with Data response, or

a Network Query response.

The first three may be returned by TCP/IP commands, the latter from a UDP/IP command.

The error response consists of the letter ‘N’ (for NAK, or negative acknowledge), followed by a 2-digit hexadecimal error code. The following table lists the error codes that can be returned from a Model 9116 module:

 

Table 3.1

 

Error Codes

 

 

CODE

MEANING

 

 

00

(Unused)

 

 

01

Undefined Command Received

 

 

02

Unused by TCP/IP

 

 

03

Input Buffer Overrun

 

 

04

Invalid ASCII Character Received

 

 

05

Data Field Error

 

 

06

Unused by TCP/IP

 

 

07

Specified Limits Invalid

 

 

08

NetScanner error; invalid parameter

 

 

09

Insufficient source air to shift calibration valve

 

 

0A

Calibration valve not in requested position

 

 

The Acknowledge response is returned from a module when a command is received that requires no data to be returned, and no error is detected. It indicates successful parsing and execution of the last received command. It consists of the letter ‘A’ (for ACK, or acknowledge).

The Acknowledge with Data response is returned when a module receives a command requesting data. Model 9116 modules will typically return only the requested data values, each preceded by a space character (except for format 7). No ‘A’ acknowledge letter begins this data response. Data are returned for the highest requested channel number first.

Page 22

www.PressureSystems.com

Page 31
Image 31