Secure the PC-323 board in place using the long screw through the mounting bar and the short screw into the hex standoff.

Figure 5.3c

PC-322 Board

(4)Turn the assembly over. Replace the nylon washer and secure the nut to the back of the long screw going through the mounting bar as depicted in Figure 5.3c.

(5)Reassemble the scanner as previously described and test for proper operation.

5.1.4Replacement of Transducers

Your Model 9116 has internal DH200 pneumatic transducers, as well as an internal calibration manifold with associated valves and O-rings. All these elements occasionally require service or replacement as described in the following sections.

Following is a step-by-step procedure to replace a DH200 transducer in a Model 9116 Intelligent Pressure Scanner. Use the tools and follow the general warnings already described in Section 5.1.1.

Figure 5.4

Top View of DH200

Page 109

www.PressureSystems.com

Page 118
Image 118