Pressure Systems, Inc.

Model 9116 User’s Manual

 

 

Internal firmware limits calculated gains to values are software limited to values between 0.0 and 100.0. Any calculated value outside of this range will result in the gain coefficient being set to 1.00.

Note

Example:

The calculated gain values from the latest ‘Z’ command will be lost when the module is powered off. To save these gain terms to each transducer’s non-volatile memory, refer to the Set Operating Options (‘w’) command (index 09).

Send TCP/IP command to a Model 9116 module (via its open socket) to calculate and set gain coefficients for channels 8 through 4. Instruct the module to use 14.8890 psi as the applied pressure instead of each transducer’s full-scale value:

Z00F8 14.8890

Read response, containing the new gain values (also stored in the module’s volatile main memory):

1.000212 1.000269 1.000437 1.000145 .999670

Actual gain values are returned in the above response as decimal ASCII strings, each preceded by a space character. From left-to-right: they are for channels 8, 7, 6, 5, and 4.

Page 42

www.PressureSystems.com

Page 51
Image 51
Measurement Specialties 9116 user manual Z00F8, 000212 1.000269 1.000437 1.000145