Pressure Systems, Inc.

Model 9116 User’s Manual

 

 

Note

With the exception of binary formats 7 and 8, all other formats include a leading space in each datum delivered in each stream packet.

The number of stream packets (‘num’) parameter is a positive integer count (from 0 to 2147483647, specified with 1 to 10 numeric digits as needed). It sets a finite limit on the number of packets delivered in the host data stream. The value 0 for this parameter requests “continuous” output packets for the defined host stream (unbounded).

Note

While the ‘c 00’ command will allow hardware and software trigger scan lists to operate concurrently, this is not a recommended mode of operation. Running hardware and software trigger scan lists concurrently diminishes the degree of determinism for the hardware scan lists. The hardware trigger lists will determine the frequency of pressure readings and the update of internal pressure, current value table. The software trigger lists will return data from this table at the requested software trigger rate, however the table will only be updated at the rate of the fastest hardware trigger scan lists.

Page 50

www.PressureSystems.com

Page 59