Measurement Specialties 9116 Calculate and SET Offsets Command ‘h’, Hpppp, Gggg .. g.gggg

Models: 9116

1 143
Download 143 pages 39.58 Kb
Page 72
Image 72

Pressure Systems, Inc.

Model 9116 User’s Manual

 

 

CALCULATE AND SET OFFSETS (Command ‘h’)

Purpose: Instructs a module to calculate new offset coefficients with zero differential pressure (or a specified “generated” pressure) applied to the specified channels. These new coefficients update part of the module’s internal calibration coefficient database, used to convert any subsequent raw data into engineering units data. The new offset values are also returned in the response. This command is sometimes called a Re-zero or zero-only calibration.

Command

hpppp [vv.vvvv]

 

h’ is the command letter

 

pppp’ is the position field

 

vv.vvvvis an [optional] applied pressure value preceded by a space

 

character

 

 

Response

g.gggg .. [g.gggg]...”

g.gggg’ are the actual offset data values returned, each preceded by a

 

 

space.

Description: The position field may have 0 or 4 characters. If no position field is specified, offset coefficients for all of a module’s input channels will be calculated and returned. If a position field is specified, offset coefficients for only the channels whose bits are set (=1) will be calculated and returned. If the optional pressure value [ vv.vvvv] is specified, the position field must be 4 characters, even when all channels are to be specified. If the optional pressure value [‘vv.vvv’] is not provided, an applied pressure of 0.0 psi(a) will be assumed when calculating coefficients. Offset values are returned in the response in order of highest specified channel to lowest specified channel, with data formatted per an implied decimal format (f=0).

Before acquiring data with this command, any addressed Model 9116 module will normally attempt to place the calibration valve in the CAL position, so that a zero differential pressure can be applied to all channels via the module’s CAL and CAL Ref input port. Simply leaving these ports unattached will allow the transducers to read the appropriate zero differential pressure if ambient air pressure is stable. After data are acquired, the calibration valve will be placed in the RUN position. To disable the automatic shifting of the calibration valve, refer to the Set Operating Options (‘w’) command (index 0B). The reader is also referred to Section 4.2 of Chapter 4 for additional details concerning the performance of a Re-zero Calibration.

Note

The calculated offset values from the latest ‘h’ command will be lost when the module is powered off. To save these offset terms to each transducer’s non-volatile memory refer to the Set Operating Options (‘w’) command (index 08).

Page 63

www.PressureSystems.com

Page 72
Image 72
Measurement Specialties 9116 user manual Calculate and SET Offsets Command ‘h’, Hpppp, Gggg .. g.gggg