Measurement Specialties 9116 Command ‘c’- Sub-command Index 03 Clear Streams, Command 03 st

Models: 9116

1 143
Download 143 pages 39.58 Kb
Page 64
Image 64

Pressure Systems, Inc.

Model 9116 User’s Manual

 

 

Command ‘c’— Sub-command Index 03: Clear Stream(s)

This sub-command is used to “undefine” any previously configured host stream in a module, one at a time, or all together. The sub-command’s format is:

Command

c 03 st

 

 

‘c’ is the command letter

 

‘03’ is the sub-command index (‘ii’) for configuration

 

st’ is the stream identifier character (1, 2, or 3 or 0=all streams)

 

NOTE: all parameters are separated by a space character.

 

 

Response

A

‘A’ is the acknowledge letter

 

 

 

Autonomous

none generated

Packet

 

 

 

Description: This sub-command clears (un-defines) the particular specified host stream (st=1- 3), or un-defines “all configured” host streams with a single command (st=0). Once cleared, a stream must be reconfigured before it can be started.

Example:

Stop all the streams configured previously. Then clear (un-define) only stream 3. Finally, resume the remaining defined streams 1 and 2:

c 02 0” “c 03 3” “c 01 0