Pressure Systems, Inc.Model 9116 User’s Manual

bbbb (hex)

data selected for inclusion in each stream packet

 

 

0001 **

Enable Valve Position Status (reserved for future use)

 

 

0002

Enable DH Temperature Status (see bit map below)

 

 

0010

Enable Pressure EU Data (default if “05” never executed after “00”)

 

 

0020

Enable Pressure A/D Counts

 

 

0040

Enable Pressure Voltages

 

 

0080

Enable DH Temperature EU Data (degrees C)

 

 

0100

Enable DH Temperature A/D Counts

 

 

0200

Enable DH Temperature Voltages

 

 

Any DH Temperature Status datum is delivered as a two-byte binary bit map (16-bit, big endian) with each bit representing the status of DH #16 through DH #1 respectively. A bit value of 0 (zero) indicates the DH is operating within its specified limits. A value of 1 (one) indicates the DH is outside its specified limits.

Bit #

Chan # Binary

Hex

15

14

13

12

11

10

9

8

7

6

5

4

3

2

1

0

16

15

14

13

12

11

10

9

8

7

6

5

4

3

2

1

1

0

 

0

0

0

0

 

0

0

0

0

 

0

0

0

0

 

0

1

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

8

 

 

 

0

 

 

 

0

 

 

 

1

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

The above example indicates that Channels 1 and 16 are operating outside the specified temperature limits.

Note

**This status field (0001) cannot be specified for Model 9116. However it is shown should the capability be added to future firmware versions. Currently, only Models 9816 and 903x can return Valve Position status in their streams.