Measurement Specialties 9116 user manual V00800-01 0.000 1.000 Response returned is a

Models: 9116

1 143
Download 143 pages 39.58 Kb
Page 91
Image 91

Pressure Systems, Inc.

Model 9116 User’s Manual

 

 

Example:

Send TCP/IP command to module (i.e., via its connected socket): with replacement values for the channel’s offset and gain correction terms loaded into the module’s volatile memory (cc=00-01). Load these into channel # 8's Transducer Coefficient array (aa=08).

v00800-01 0.000 1.000” Response returned is: “A

Send command to Model 9116 module (via its connected socket) to change its default EU output from psi to kPa. This will be done by changing the EU Pressure Conversion Scaler to 6.894757.

v01101 6.894757” Response returned is “A

Page 82

www.PressureSystems.com

Page 91
Image 91
Measurement Specialties 9116 V00800-01 0.000 1.000 Response returned is a, V01101 6.894757 Response returned is a