Measurement Specialties 9116 Read Module Status Command ‘q’, Firmware Version, as hex value

Models: 9116

1 143
Download 143 pages 39.58 Kb
Page 78
Image 78

Pressure Systems, Inc.

Model 9116 User’s Manual

 

 

READ MODULE STATUS (Command ‘q’)

Purpose: Returns requested module status information.

Command

qii

 

 

q’ is the command letter

 

ii’ is the status index field

Response

hhhh

hhhh’ is a 4-digit hex datum (or other (**) decimal datum

 

Description: The 2-digit hex index field (ii) chooses a particular status field to be returned. Returned value is described in the following table for each index (a third column shows any ‘w’ command index for setting same option.

 

ii

 

 

returned value

 

 

‘w’ set

 

 

 

 

4-digit hex or other decimal (**)

 

 

index

 

 

 

 

 

 

 

 

00

 

 

Module’s Model Number, as decimal (**) integer value (e.g, 9116).

 

 

 

 

 

 

 

 

 

 

01

 

 

Firmware Version, as hex value

 

 

 

 

 

 

 

(expressed internally as integer version * 100).

 

 

 

 

 

 

 

(e.g. hex ‘0100’ = 256 decimal, means Version 2.56)

 

 

 

02

 

 

Power-up Status, as 16-bit hex bit map, bits having the following

 

 

 

 

 

 

 

meaning:

 

 

 

 

 

 

 

Bit 0 (LSB): A/D Failure Error.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Bit 1: Transducer Re-zero Adjustment (offset) Term Range Error (out-

 

 

 

 

 

 

 

of-range values set to 0.0 internally).

 

 

 

 

 

 

 

Bit 2: Transducer Span Adjustment (gain) Term Range Error (out-of-

 

 

 

 

 

 

 

range values set to 1.0 internally).

 

 

 

 

 

 

 

Bit 3: Temperature Correction Coefficients Not Present or Out-of-Range

 

 

 

 

 

 

 

(if transducer has one or more bad coefficients, all set to 0.0).

 

 

 

 

 

 

 

Bit 4: reserved (for transducer checksum)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Bit 5: FLASH Initialized Data Section Checksum Error (if error, all data

 

 

 

 

 

 

 

variables set to factory defaults and stored in FLASH).

 

 

 

 

 

 

 

Bit 6: SRAM Error.

 

 

 

 

 

 

 

 

 

03

 

reserved

 

 

 

 

 

 

 

 

 

04

 

reserved

 

 

 

 

 

 

 

 

 

 

 

05

 

 

Number of A/D Samples To Average, as hex value (e.g., 000A=10

10

 

 

 

decimal).

 

 

 

 

 

 

 

 

Page 69

www.PressureSystems.com

Page 78
Image 78
Measurement Specialties 9116 Read Module Status Command ‘q’, Returned value ‘w’ set Digit hex or other decimal Index