Pressure Systems, Inc.Model 9116 User’s Manual

 

cc

 

 

Transducer Coefficients Description

 

 

Datum

 

 

 

 

 

 

Type

 

 

 

 

 

 

 

 

 

 

00

 

 

Re-zero Cal Adjustment (offset) term

(Note 1)

 

FLOAT

 

 

 

 

 

 

 

01

 

 

Span Cal Adjustment (gain) term

(Note 2)

 

FLOAT

 

 

 

 

 

 

 

07

 

 

User Defined Field

(Note 4)

 

INTEGER

 

 

 

 

 

09

 

Transducer Manufacturing Reference Number (Note 5)

 

INTEGER

 

 

 

 

 

 

 

0A

 

Transducer Full-Scale Range Code (See Appendix F) (Note 5)

 

INTEGER

 

 

 

 

 

 

 

 

 

 

(Note 1) Related command ‘w08’ can be used to download the offset term to the sensor’s non-volatile memory (digitally or non-digitally compensated sensor).

(Note 2) Related command ‘w09’ can be used to download the gain term to the sensor’s non-volatile memory (digitally or non-digitally compensated sensor).

(Note 4) Data is immediately stored to the sensor’s non-volatile memory.

The User Defined Date field (cc=07) is a 32-bit integer. Possible uses are to I ndicate the date of last user zero and/or span calibration or possibly the date of next required calibration. If this optional field is used, the user is responsible for correctly encoding the date into the appropriate 32-bit integer value e.g., a decimally encoded ‘yymmdd’ date. Any modifications of this field (using the Download Internal Coefficients (‘v’) command) will result in the new value automatically being entered to transducer non-volatile memory.

A special single Other Coefficients array is selected with array index aa=11 (hex). All the valid coefficient indexes (for this array only) are listed in the following table:

cc

Other Coefficients Description

 

 

01

EU Pressure Conversion scaler (default=1.0)

 

 

Datum

Type

FLOAT

Page 81

www.PressureSystems.com

Page 90
Image 90
Measurement Specialties 9116 user manual Other Coefficients Description, Datum Type