channels. Following completion of the acquisition and EU conversion cycle, the module will also deliver the requested data channels to the host. In this manner, users are provided with highly synchronized data acquisition and delivery from one or more modules. If a module waits in the idle mode for an extended period of time without receiving a data request, it will periodically initiate its own internal data acquisition cycles so as to update internal thermal coefficients.

When all hardware triggered streams are complete or aborted, an individual module will return to the default mode of continuous scanning and EU conversion.

When the internal software timer is used to control host stream output rates (sync=1), note that internal clock frequency variances will result in slightly different timing between modules. Although these differences in timing are slight, they may result in noticeable differences in output timing between modules over a long period of time. If highly synchronized data output is required from multiple modules, the hardware trigger mode should be used.

The period ‘per’ parameter is a positive decimal integer count (from 0 to 2147483647, specified with 1 to 10 numeric digits as needed), and its meaning depends on the sync type sync’ parameter described above.

‘sync’

meaning of ‘per’

 

 

0number of hardware trigger periods to wait before sending each packet delay period (in milliseconds) to wait before sending each packet

1NOTE: minimum is 2 milliseconds and the granularity is 2 milliseconds with values rounded down to the nearest 2 milliseconds.

The ‘f’ parameter identifies the format of each selected acquired datum in each stream packet, and is a single numeric digit. Valid format codes are listed in the following table:

f

converts each internal selected acquired datum value from . . .

max. char.

 

 

 

 

 

0

single binary float

to

7-10-digit signed decimal “ [-xxx]x.xxxxxx”

13

 

 

 

 

 

1

single binary float

to

8-digit hex integer “ xxxxxxxx”

9

 

 

 

 

 

2

double binary float

to

16-digit hex integer “ xxxxxxxxxxxxxxxx”

17

 

 

 

 

 

5

single binary float

to

long integer (EU*1000) then to 8-digit hex integer

9

 

 

 

 

 

7

single binary float

to

single binary float (big endian: msb first)

4

 

 

 

 

 

8

single binary float

to

single binary float (little endian: lsb first)

4

 

 

 

 

 

Unless the EU conversion scalar is altered, the returned pressure data will be in units of psi.

(See command ‘v’, array 11, coefficient 01 for other units.