Pressure Systems, Inc.

Model 9116 User’s Manual

 

 

5.1.6.5Solenoid Valve O-Ring Replacement

Following is a step-by-step procedure to replace the internal solenoid valve O-rings in a Model 9116 Intelligent Pressure Scanner. The module contains two internal solenoid valves.

(1) Disassemble the module as described in Section 5.1.2.

(2)Remove the solenoid valve by unscrewing the two (2) Phillips-head screws on top of the solenoid. Gently lift it out of the module. Be careful not to crimp the attached nylon tubing.

(3)Remove and replace the O-rings needing maintenance using the procedure described in Section 5.1.6.1.

(4)Replace the solenoid valve and gently tighten the screws.

(5)Reassemble the module.

Figure 5.7

Solenoid Valve O-Ring Replacement

(6) Test your scanner to ensure proper operation.

Page 116

www.PressureSystems.com

Page 125
Image 125
Measurement Specialties 9116 user manual Solenoid Valve O-Ring Replacement