Example:

Supply each of the previously-specified three (3) pressure calibration points to the Multi- Point Calibration function, as was stated in the previous example of the Configure and Start (‘00’) sub-command. Assume that all the affected four (4) channels have differential transducers with the same -5 to +5 psi range. Include at least one pressure point in the negative range of these transducers

C 01 1 -2.5C 01 2 0.0” “C 01 3 5.0

Read responses (separately after each command executed above):

“-2.4998 -2.4999 -2.5001 -2.500” “0.0 0.0013 -0.0133 -0.00001” “5.0091 4.9992 5.0010 4.9998”

Data are returned in reverse channel number order (i.e., 4, 3, 2, 1) in each response.

Page 35

www.PressureSystems.com

Page 44
Image 44