Model 9116 User’s Manual

 

 

Following is a simple, step-by-step procedure for executing a specified-value span calibration of a Model 9116 Intelligent Pressure Scanner. For the purposes of this example, it will be assumed that an upscale pressure of 14.9800 psi is available from a dead weight tester for the calibration of 15 psi internal transducers. All sixteen channels are 15 psi full scale.

 

Description

 

Model 9116 Command

 

 

 

 

 

Ensure that valves are in RUN/CAL mode (default)

 

[w1200]

 

 

 

 

 

… normal data acquisition

 

 

 

 

 

 

 

Perform Re-zero calibration first

 

See Section 4.2.2

 

 

 

 

 

Place the module calibration manifolds into the CAL position

 

[w0C01] for CAL

 

if this is the desired pressure application input. The RUN

 

 

 

 

 

position may be a better choice for modules with

 

[w0C00] for RUN

 

transducers having different ranges.

 

 

 

 

 

With a deadweight tester, apply 14.9800psi to the

 

 

 

appropriate module CAL and CAL REF inputs [or optionally

 

 

 

to RUN inputs]

 

 

 

Delay for settling of pneumatic inputs

 

 

 

 

 

 

 

Verify measured data reads near expected full scale

 

[rFFFF0]

 

 

 

 

 

Instruct module to calculate new gain coefficients for all 16

 

ZFFFF 14.98

 

channels

 

 

 

 

 

Place calibration manifold back into the RUN position

 

w0C00

 

 

 

 

 

Store new gain coefficients to transducer non-volatile

 

w09

 

memory

 

 

 

 

 

… continue normal data acquisition

 

 

 

 

 

 

Page 97

www.PressureSystems.com

Page 106