Pressure Systems, Inc.Model 9116 User’s Manual

 

IP Address Resolution Method, as hex state: (default = 0000)

 

 

0000 = Use Static IP Address stored in module’s non-volatile

 

06

memory

13

 

0001 = Get Dynamic IP Address from external RARP/BOOTP

 

 

server

 

 

Host Response/Stream Back-Off Delay, as hex value (or FFFF). FFFF

 

07

means use low-order byte of module’s Ethernet Address as value

14

instead. In either case, back-off delay in microseconds is calculated from

 

decimal equivalent of hex value:

 

 

delay = decvalue * 20

 

 

Host Response/Stream Total Size Prefix (with 2-bytebig-endian binary

 

 

value), added to all command responses and streams to indicate their

 

08

true length in bytes:

16

 

0000 = None (default)

 

 

0001 = Yes

 

09

TCP Connect Port, as hex value (e.g. 2328 = 9000 decimal, default).

17

 

 

 

 

Auto UDP Broadcast@Reset, as hex state:

 

0A

0000 = No (default)

18

 

0001 = Yes

 

 

Temperature Status of Each Scanner Transducer, as 16-bit hex bit

 

 

map, each bit representing the current status of a transducer/channel

 

0C

(16-1)). Bit values are:

 

0= transducer operating within the specified operational limits.

 

 

 

 

1= transducer operating outside the specified limits.

 

 

(see end-of-table NOTE +)

 

0D

Minimum Temperature Alarm Set Point (in degrees C), as decimal (**)

19

format 0 representation of internal IEEE float, with leading space).

 

 

0E

Maximum Temperature Alarm Set Point (in degrees C), as decimal (**)

19

format 0 representation of internal IEEE float, with leading space).

 

 

11

Thermal Update Scan Interval (in seconds) as decimal (**) integer

1B

value.

 

 

31

Module hardware version number in the form e.eeeeee. (9016 returns

 

‘N08’

 

 

 

 

Hardware trigger mode.

 

32

0=positive going edge

32

1=negative going edge

 

2=trigger on any edge

 

 

(9016 returns a ‘N08’)

 

 

Temperature Range, as a hex value

 

3c

0000 = range 0 to 60ºC (default)

 

0006 = range -30 to 60º C

 

 

 

 

0007 = range -20 to 70ºC

 

(+)NOTE: This 4-byte hex status fields may also be returned in autonomous data streams, but as pure binary extensions of each stream’s packet binary header (see ‘c’ command, ii=05, bbbb=0002).