Pressure Systems, Inc.Model 9116 User’s Manual

Verify that pressure reads correctly.

[rFFFF0]

 

 

When the data are stable, enter another Collect Data

C 01 2 5.0

(‘01’) sub-command of ‘C’ specifying this second

 

calibration point (2) with 5.0 psi pressure.

 

Apply 3rd calibration pressure to the module’s CAL or

 

RUN inputs. A mid-scale negative (-2.5 psi) point is

 

assumed in this case. A vacuum pump is normally

 

required to achieve such a pressure with 903x calibrators.

 

Verify that measured pressure reads correctly.

[rFFFF0]

 

 

When the data are stable, enter last Collect Data (‘01’)

C 01 3 -2.5

sub-command of ‘C’ for this point (3) with a negative

 

(-2.5 psi) pressure.

 

Now that data have been collected for every point

C 02

originally specified, calculate and apply the new coefficient

 

data with a Calculate and Apply (‘02’) sub-command of

 

C’. This also restores the module to using its original

 

averaging parameters that existed before the first ‘C

 

command.

 

Place calibration manifold back into the RUN position, if

[w0C00]

the CAL position was used.

 

Store new offset and gain coefficients into transducer non-

w08

volatile memory.

w09

… continue normal data acquisition.

 

 

 

4.5Coefficient Storage

The various calibration functions described in Sections 4.2 through 4.4 update the active offset and gain coefficients, respectively, in the module’s volatile main memory (RAM) only. These newer calibration coefficients will be lost when instrument power is turned off. The Set/Do Operating Options (‘w’) command may be used to also store these coefficients back in each transducer’s nonvolatile memory. This command’s option index = 08 will store new offset coefficients, while its option index = 09 will store new gain coefficients.

A user may read (and should verify) any new offset and/or gain coefficients after performing each calibration adjustment command (i.e., by saving coefficient data returned in a command ‘h’ or ‘Z’ response), or the Read Internal Coefficients (‘u’) command may be used to read them any time after calibration adjustment commands have been performed (see coefficient indexes cc=00 and 01 for arrays aa=01 through 10). These “adjusted” coefficients may be verified, and then saved by storing them in each transducer’s non-volatile memory with the ‘w’ command described above.

Page 100

www.PressureSystems.com

Page 109
Image 109
Measurement Specialties 9116 user manual Coefficient Storage