Pressure Systems, Inc.Model 9116 User’s Manual

 

ii

 

 

dd

 

 

Description

 

 

‘q’ read

 

 

 

 

 

 

 

 

index

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

00

 

 

Set Cal Valves to RUN or LEAK Position (default)

 

0B

 

0C

 

 

 

 

— choice made by ii=12.

 

 

 

 

 

 

 

see chart

 

01

 

 

Set Cal Valves to CAL/RE-ZERO or PURGE Position

 

 

 

 

 

 

 

below

 

 

 

 

 

 

 

choice made by ii=12.

 

 

 

 

 

 

 

 

 

 

 

 

0D-0E

--

 

 

Reserved for factory use

 

 

 

 

 

 

 

 

 

 

 

 

 

 

0F

00

 

 

Disable periodic Thermal Coefficient Update task.

 

 

 

 

01

 

 

Enable periodic Thermal Coefficient Update task (default).

 

 

 

 

 

 

 

 

 

 

 

 

 

 

01-20

 

Set Number of A/D Samples to Average, (default = 8).

05

 

10

 

 

 

 

 

Valid values are 4, 8,16, 32, and 64. Other values below 64

 

 

 

 

 

 

 

 

 

 

are rounded up to the next valid value listed above.

 

 

 

11

 

--

 

 

Reserved for factory use

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

00

 

 

Set Cal Valves to RUN or CAL/RE-ZERO Position (default)

 

0B

12

 

 

 

 

 

— choice made by ii=0C.

 

 

 

 

 

 

 

see chart

 

01

 

 

Set Cal Valves to PURGE or LEAK Position — choice made

 

 

 

 

 

 

 

 

by ii=0C.

 

below

 

 

 

 

 

 

 

 

 

 

 

 

 

00

 

 

Use Static IP Address Resolution (default)

06

 

13

 

01

 

 

Use Dynamic IP Address Resolution (RARP/BOOTP)

 

 

 

 

 

 

 

 

(Results in immediately becoming the module’s new power-on

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

default)

 

 

 

 

 

 

00

 

 

Disable Host Response/Stream Back-Off Delay (default).

07

 

 

 

 

01

 

 

Enable Host Response/Stream Back-Off Delay as low-order

 

 

 

14

 

 

 

 

 

byte of Ethernet Address(converted to decimal value * 20

 

 

 

 

 

 

 

 

￿sec.).

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

02

 

 

Enable Host Response/Stream Back-Off Delay specified

 

 

 

 

 

 

 

 

 

 

per eeee as decimal value (* 20 ￿sec.)

 

 

 

 

 

 

00

 

 

Disable Host Response/Stream Total Size Prefix (default).

08

 

16

 

 

 

 

 

Enable Host Response/Stream Total Size Prefix (2-byte

 

 

 

 

01

 

 

big-endian binary value with total size of response or stream

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

data in bytes that follows it).

 

 

 

17

 

00

 

Set TCP Connect Port per eeee as decimal value

09

 

 

 

 

 

 

(default=9000).

 

 

 

 

 

 

 

 

 

 

 

 

 

18

 

00

 

 

Disable Auto UDP Broadcast at Reset (default).

 

0A

 

01

 

 

Enable Auto UDP Broadcast at Reset.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

00

 

 

Set Minimum Temperature Alarm Set Point (in NC)

 

0D

19

 

 

 

 

 

per eeee as decimal value (default = 0 NC).

 

 

 

 

01

 

 

Set Maximum Temperature Alarm Set Point ( in NC)

 

0E

 

 

 

 

 

 

 

per eeee as decimal value (default = 60 NC).

 

 

 

 

1B

00

 

Set Thermal Update Scan Interval per eeee as decimal

11

 

 

 

 

 

 

value (seconds), 1 <= eeee <= 3600 seconds (default = 15).

 

 

 

 

 

 

 

 

 

 

 